PEL-3000(H) User Manual
174
Note
Use resistors with minimum residual resistance of
50
Ω
or less.
Note for proportional control: Do not use swtiches
that switch between fixed resistances. Please use
continuously variable resistors.
External Resistance Control – Operation
Description
External resistance control can be used to control
the current, voltage, resistance and power for CC,
CR, CV and CP modes. Configuration for each
operating mode is the same.
CC Mode
Proportional Control:
Input current = rated current × (external
resistance/10).
Inverse Control:
Input current = rated current × (1 - external
resistance/10).
Input
Current
External
Resistance
Rated
Current
0
Ω
10k
Ω
Proportional control
Inverse control
Summary of Contents for PEL-3000 Series
Page 1: ...DC Electronic Load PEL 3000 H Series USER MANUAL VERSION 2 00 ISO 9001 CERTIFIED MANUFACTURER ...
Page 24: ...PEL 3000 H User Manual 22 PEL 3111H PEL 3211 Booster Pack PEL 3211H Booster Pack ...
Page 41: ...GETTING STARTED 39 Fasteners 2 ...
Page 207: ...REMOTE CONTROL 205 6 7 8 9 The function check is complete ...
Page 229: ...APPENDIX 227 Operating Area PEL 3021 ...
Page 230: ...PEL 3000 H User Manual 228 PEL 3041 ...
Page 231: ...APPENDIX 229 PEL 3111 ...
Page 232: ...PEL 3000 H User Manual 230 PEL 3211 Booster Pack PEL 3021H ...
Page 233: ...APPENDIX 231 PEL 3041H ...
Page 234: ...PEL 3000 H User Manual 232 PEL 3111H PEL 3211H Booster Pack ...
Page 235: ...APPENDIX 233 ...