PEL-3000(H) User Manual
188
Parallel Operation
The PEL-3000(H) series can be connected in parallel to increase the
total power capacity of a single unit.
The PEL-3000(H) series can operate with up to 5 units in parallel. A
single unit is designated as a master unit and any other connected
units as slaves.
Only units of the same type and rating can be used in parallel or
alternatively, the PEL-3211(H) booster pack can be used as a slave
with the PEL-3111(H).
When a master unit is used in parallel mode, to ensure stability, the
response speed will drop down to 1/2 if it was originally 1/1. You
can however, reset the response speed back (or to another value) in
the Main>Configure menu.
Parallel Capacity, PEL-3021, PEL-3041, PEL-3111
Model
Single Unit 2 Units
3 Units
4 Units
5 Units
PEL-3021
150V
35A
175W
150V
70A
350W
150V
105A
525W
150V
140A
700W
150V
175A
875W
PEL-3041
150V
70A
350W
150V
140A
700W
150V
210A
1050W
150V
280A
1400W
150V
350A
1750W
PEL-3111
150V
210A
1050W
150V
420A
2100W
150V
630A
3150W
150V
1680A
4200W
150V
1050A
5250W
Summary of Contents for PEL-3000 Series
Page 1: ...DC Electronic Load PEL 3000 H Series USER MANUAL VERSION 2 00 ISO 9001 CERTIFIED MANUFACTURER ...
Page 24: ...PEL 3000 H User Manual 22 PEL 3111H PEL 3211 Booster Pack PEL 3211H Booster Pack ...
Page 41: ...GETTING STARTED 39 Fasteners 2 ...
Page 207: ...REMOTE CONTROL 205 6 7 8 9 The function check is complete ...
Page 229: ...APPENDIX 227 Operating Area PEL 3021 ...
Page 230: ...PEL 3000 H User Manual 228 PEL 3041 ...
Page 231: ...APPENDIX 229 PEL 3111 ...
Page 232: ...PEL 3000 H User Manual 230 PEL 3211 Booster Pack PEL 3021H ...
Page 233: ...APPENDIX 231 PEL 3041H ...
Page 234: ...PEL 3000 H User Manual 232 PEL 3111H PEL 3211H Booster Pack ...
Page 235: ...APPENDIX 233 ...