PEL-3000(H) User Manual
222
Operating Mode Description
CC Mode
CC Mode
When the unit is set to CC mode it will operate as a
constant current load when connected to a
constant voltage source. This means the unit will
sink a designated amount of current, up to the
rated power level, regardless of the voltage. This is
illustrated below.
CC Mode
Load Input
Voltage
Load Current
CC Current
Value
CC+CV Mode
When CC+CV mode is enabled, the unit will act as
constant current load after the input voltage is
greater than the user-defined CV level. At the CV
level, the unit works as a constant voltage load.
This mode effectively creates a voltage ceiling
before the unit operates in CC mode. The diagram
below illustrates this.
Summary of Contents for PEL-3000 Series
Page 1: ...DC Electronic Load PEL 3000 H Series USER MANUAL VERSION 2 00 ISO 9001 CERTIFIED MANUFACTURER ...
Page 24: ...PEL 3000 H User Manual 22 PEL 3111H PEL 3211 Booster Pack PEL 3211H Booster Pack ...
Page 41: ...GETTING STARTED 39 Fasteners 2 ...
Page 207: ...REMOTE CONTROL 205 6 7 8 9 The function check is complete ...
Page 229: ...APPENDIX 227 Operating Area PEL 3021 ...
Page 230: ...PEL 3000 H User Manual 228 PEL 3041 ...
Page 231: ...APPENDIX 229 PEL 3111 ...
Page 232: ...PEL 3000 H User Manual 230 PEL 3211 Booster Pack PEL 3021H ...
Page 233: ...APPENDIX 231 PEL 3041H ...
Page 234: ...PEL 3000 H User Manual 232 PEL 3111H PEL 3211H Booster Pack ...
Page 235: ...APPENDIX 233 ...