PEL-3000(H) User Manual
76
Display
0.000
V
0.0000
A
0.00
W
Level1
0.00
LOAD
RS232
01/Oct/2017
W
0:00:05
Elapsed time
Cut Off Time
Description
The Cut Off Time function will turn the load off
after a set-amount of time. After the load has been
turned off, a popup screen will display the voltage
level when the load was turned off.
Operation
1.
Press
Main
>
Configure[F5]
>
Other[F2]
and
set the
Cut Off Time
.
Range:
OFF, 1 second - 999 hours:59
minutes:59 seconds
Display
Mode
CP
I Range
H 8.75A
V Range
L 80V
Function
Dynamic
Configure
0.000
V
0.0000
A
0.00
W
Timer1
0.025
Level2
0.00
Level1
0.00
LOAD
RS232
Fine
01/Oct/2018
mS
W
W
0:00:05
Cut off time
Time Up
Voltage : 5.1223V
Enter
Voltage at cut off time