A COM
19 Connected to the negative load input terminal on
the rear panel.
+15V
20 Controls the on/off of the load booster power
(cannot be used for multiple purposes).
*1 Valid only when the front panel settings are
H range.
*2
RANGE CONT 0
H range 1
M range 1
*3 The maximum applied voltage of the photocoupler
is 30 V; the maximum current is 8 mA.
J2 Connector Booster
Pin name
Pin number
Description
N.C.
1 Not connected.
N.C.
2 Not connected.
N.C.
3 Not connected.
SUM I MON
4 Connect to SUM I MON of the J1 connector.
PRL OUT+
5 Used during master/slave operation. Connected to
PRL IN+ of the J1 connector.
PRL OUT-
6 Used during master/slave operation. Connected to
PRL IN- of the J1 connector.
LOAD ON/OFF
CONT
7
N.C.
8 Not connected.
SLAVE RANGE
CONT
9 Used during master/slave operation. Connected to
RANGE CONT 0 of the J1 connector.
N.C.
10 Not connected.
N.C.
11 Not connected.
A COM
12 Connected to the negative load input terminal on
the rear panel.
N.C.
13 Not connected.
N.C.
14 Not connected.
N.C.
15 Not connected.
ALARM INPUT
16 Activates an alarm with high (or low) TTL level
signal input. Pulled up by the internal circuit to 5 V.
Summary of Contents for PEL-3000 Series
Page 1: ...DC Electronic Load PEL 3000 H Series USER MANUAL VERSION 2 00 ISO 9001 CERTIFIED MANUFACTURER ...
Page 24: ...PEL 3000 H User Manual 22 PEL 3111H PEL 3211 Booster Pack PEL 3211H Booster Pack ...
Page 41: ...GETTING STARTED 39 Fasteners 2 ...
Page 207: ...REMOTE CONTROL 205 6 7 8 9 The function check is complete ...
Page 229: ...APPENDIX 227 Operating Area PEL 3021 ...
Page 230: ...PEL 3000 H User Manual 228 PEL 3041 ...
Page 231: ...APPENDIX 229 PEL 3111 ...
Page 232: ...PEL 3000 H User Manual 230 PEL 3211 Booster Pack PEL 3021H ...
Page 233: ...APPENDIX 231 PEL 3041H ...
Page 234: ...PEL 3000 H User Manual 232 PEL 3111H PEL 3211H Booster Pack ...
Page 235: ...APPENDIX 233 ...