OPERATION
71
Display
Mode
CV
I Range
H 8.75A
V Range
L 80V
Response
Slow
Configure
0.000
V
0.0000
A
0.00
W
CV B Value
80.000
CV A Value
80.000
LOAD
RS232
A Value
Fine
01/Oct/2018
V
V
Response
setting
CC, CR and CP Mode Response Speed
Description
By default, the “normal current response” speed is
set to 1/1. The response speed can be reduced to
1/2, 1/5, 1/10.
Reducing the current response speed can affect
other settings such as the slew rate and soft
start settings.
Operation
1.
Make sure the load is off.
2.
Press
Main
>
Configure[F5]
>
Other[F2]
and
set the
Response
parameter.
Range:
1
1
,
2
1
,
5
1
,
10
1