PEL-3000(H) User Manual
194
Turning the Load On
Description
Operating the PEL-3000(H) Series in parallel mode
is the same as for single units.
Note
When using the units in parallel, the load line
inductance could be increased or the stability of the
units could be reduced. It may be necessary to reduce
the response speed setting to increase stability.
1.
Turn the slave and master units on.
2.
Set the operation mode and settings on the
master unit.
The master’s settings will be used by the slave
units.
3.
Turn the load on from the Master unit.
All measurements will be displayed and
updated on the Master unit only.
Disable Parallel Mode
Description
To disable parallel mode, each unit must be set as a
“Master”.
Operation
1.
Turn the power off on all the units and remove
the GTL-255 frame link cables.
2.
Turn the power back.
3.
On each unit, press
Main
>
Configure [F5]
>
Next Menu [F4]
>
Parallel[F1].
4.
Set the unit
to
Master
with the
Operation
setting.
5.
Turn the
Parallel
and
Booster
settings to
Off
.
Summary of Contents for PEL-3000 Series
Page 1: ...DC Electronic Load PEL 3000 H Series USER MANUAL VERSION 2 00 ISO 9001 CERTIFIED MANUFACTURER ...
Page 24: ...PEL 3000 H User Manual 22 PEL 3111H PEL 3211 Booster Pack PEL 3211H Booster Pack ...
Page 41: ...GETTING STARTED 39 Fasteners 2 ...
Page 207: ...REMOTE CONTROL 205 6 7 8 9 The function check is complete ...
Page 229: ...APPENDIX 227 Operating Area PEL 3021 ...
Page 230: ...PEL 3000 H User Manual 228 PEL 3041 ...
Page 231: ...APPENDIX 229 PEL 3111 ...
Page 232: ...PEL 3000 H User Manual 230 PEL 3211 Booster Pack PEL 3021H ...
Page 233: ...APPENDIX 231 PEL 3041H ...
Page 234: ...PEL 3000 H User Manual 232 PEL 3111H PEL 3211H Booster Pack ...
Page 235: ...APPENDIX 233 ...