GETTING STARTED
31
0
8.25246
0.32242
245
1
7.34822
0.40639
211
2
6.54304
0.51266
181
3
5.82676
0.64616
158
4
5.18922
0.81508
135
5
4.62026
1.02762
118
6
4.1148
1.29593
101
7
3.66522
1.6341
89
8
3.2639
2.0605
73
9
2.90576
2.59809
64
10
2.58826
3.27639
55
11
2.30378
4.1328
47
12
2.05232
5.20864
41
13
1.8288
6.56984
35
14
1.62814
8.282
32
15
1.45034
10.44352
28
16
1.29032
13.17248
22
17
1.15062
17.60992
19
18
1.02362
20.9428
16
19
0.91186
26.40728
14
20
0.8126
33.292
11
21
0.7239
41.984
9
Load Line
Inductance
Considerations
When using the PEL-3000(H) load generator,
voltage drop and voltage generated due to load
line inductance and current change must be taken
into account. Extreme changes in voltage may
exceed the minimum or maximum voltage limits.
Exceeding the maximum voltage limit may
damage the PEL-3000(H).
To determine the voltage generated, the following
equation can be used.
E = L x (∆ I / ∆ T)
E= voltage generated
L=load line inductance
∆ I= change of current (A)
∆ T= time (us)
Load line inductance (L) can be approximated as
1uH per 1 meter of wire. (∆ I / ∆ T) is the slew rate
Summary of Contents for PEL-3000 Series
Page 1: ...DC Electronic Load PEL 3000 H Series USER MANUAL VERSION 2 00 ISO 9001 CERTIFIED MANUFACTURER ...
Page 24: ...PEL 3000 H User Manual 22 PEL 3111H PEL 3211 Booster Pack PEL 3211H Booster Pack ...
Page 41: ...GETTING STARTED 39 Fasteners 2 ...
Page 207: ...REMOTE CONTROL 205 6 7 8 9 The function check is complete ...
Page 229: ...APPENDIX 227 Operating Area PEL 3021 ...
Page 230: ...PEL 3000 H User Manual 228 PEL 3041 ...
Page 231: ...APPENDIX 229 PEL 3111 ...
Page 232: ...PEL 3000 H User Manual 230 PEL 3211 Booster Pack PEL 3021H ...
Page 233: ...APPENDIX 231 PEL 3041H ...
Page 234: ...PEL 3000 H User Manual 232 PEL 3111H PEL 3211H Booster Pack ...
Page 235: ...APPENDIX 233 ...