GE Analytical Instruments ©2009
149 of 236
DLM 90488-01 Rev. A
Figure 25: TC Values Summary Screen in a Multi-Point Calibration
Figure 26: IC Values Summary Screen in a Multi-Point Calibration
Performing a Calibration Verification
Use the verification procedure to confirm that the Analyzer’s current calibration is accurate. GE
Analytical Instruments recommends performing a verification after replacement of consumables,
such as the pump tubing, UV lamp, resin bed, or reagents, and after calibration.
Calibrate
12 Apr 2004 08:12
TC Calibration
Expected
Measured
Diff
Adjusted
Blank
N.A.
459ppb
N.A
462ppb
1ppm
1.25ppm
1.23ppm
-1.60%
1.24ppm
5ppm
5.3ppm
5.33ppm
0.57%
5.36ppm
10ppm
10.2ppm
10.3ppm
0.98%
10.4ppm
25ppm
25.2ppm
25.3ppm
0.40%
25.4ppm
50ppm
50.2ppm
49.9ppm
-0.60%
50.1ppm
R
2
1.00
Calibration passed
IC Data
Cancel
Calibrate
12 Apr 2004 08:12
IC Calibration
Expected
Measured
Diff
Adjusted
1ppm
1.52ppm
1.50ppm
-1.32%
1.51ppm
5ppm
5.70ppm
5.75ppm
0.88%
5.79ppm
10ppm
10.6ppm
10.8ppm
1.89%
10.8ppm
25ppm
25.5ppm
25.4ppm
-0.39%
25.5ppm
50ppm
50.4ppm
50.2ppm
-0.40%
50.3ppm
R
2
1.00
Calibration passed
Apply
TC Data
Cancel
Summary of Contents for Sievers 900 Series
Page 10: ...GE Analytical Instruments 2009 10 of 236 DLM 90488 01 Rev A ...
Page 88: ...GE Analytical Instruments 2009 88 of 236 DLM 90488 01 Rev A ...
Page 122: ...GE Analytical Instruments 2009 122 of 236 DLM 90488 01 Rev A ...
Page 159: ...GE Analytical Instruments 2009 159 of 236 DLM 90488 01 Rev A ...
Page 161: ...GE Analytical Instruments 2009 161 of 236 DLM 90488 01 Rev A ...
Page 165: ...GE Analytical Instruments 2009 165 of 236 DLM 90488 01 Rev A ...
Page 167: ...GE Analytical Instruments 2009 167 of 236 DLM 90488 01 Rev A ...
Page 170: ...GE Analytical Instruments 2009 170 of 236 DLM 90488 01 Rev A ...
Page 194: ...GE Analytical Instruments 2009 194 of 236 DLM 90488 01 Rev A ...
Page 218: ...GE Analytical Instruments 2009 218 of 236 DLM 90488 01 Rev A ...
Page 235: ...Notes GE Analytical Instruments 2009 235 of 236 DLM 90488 01 Rev A 236 ...
Page 236: ...Notes GE Analytical Instruments 2009 236 of 236 DLM 90488 01 Rev A 236 ...