GE Analytical Instruments ©2009
93 of 236
DLM 90488-01 Rev. A
Setting the Analyzer Mode
The Sievers 900 On-Line TOC Analyzer can operate in two different modes. To set the Analyzer
mode, follow these steps:
1. Select the
Setup
tab (see Figure 12).
• Select
On-Line
to measure TOC continuously from a sample stream or sipper tube. If a
sipper tube is used, the Flow Sensor must be disabled; see “Configuring the Flow
Sensor” on page 118 for details.
On-Line
is the default mode for the Sievers 900 On-
Line Analyzer.
• Select
Grab
to measure TOC directly from a sample flask using the sipper tube
attachment or the vial port. After you select
Grab
, you will need to press the
Configure
Mode
button to configure the settings for Grab mode. If you will be sampling from a
flask, you will also need to change the sample inlet configuration.
Grab Mode Measurement Settings
When selecting Grab mode, you need to specify the number of replicate measurements and rejects
as follows:
1. On the
Setup
tab, press the
Configure Mode
button.
2. Press the
Reps
button to set the number of replicate TOC measurements to be made on
the sample. When you press the button, a number pad displays. Enter a number and press
the
Enter
button to save the value, or press
Cancel
to retain the current setting without
making any changes.
3. Press the
Rejects
button. This option determines how many of the replicate measurements
will be rejected in calculating average values and standard deviations. When you press the
button, a number pad displays. Enter a number and press the
Enter
button to save the
value, or press
Cancel
to retain the current setting without making any changes.
4. Press the
Grab Label
button to assign an alphanumeric label of up to 7 characters to each
grab sample. The label displays in the data history. After the analysis completes, the
Grab
Label
resets to its default value.
In most cases, this first measurement will be close to the actual TOC but may be outside of the
repeatability of the Analyzer. It is best to set the number of rejections to
1
or more to ensure an
accurate TOC value. Using at least
4
repetitions allows rejection of the first value and calculation of
a standard deviation using the remaining values.
Summary of Contents for Sievers 900 Series
Page 10: ...GE Analytical Instruments 2009 10 of 236 DLM 90488 01 Rev A ...
Page 88: ...GE Analytical Instruments 2009 88 of 236 DLM 90488 01 Rev A ...
Page 122: ...GE Analytical Instruments 2009 122 of 236 DLM 90488 01 Rev A ...
Page 159: ...GE Analytical Instruments 2009 159 of 236 DLM 90488 01 Rev A ...
Page 161: ...GE Analytical Instruments 2009 161 of 236 DLM 90488 01 Rev A ...
Page 165: ...GE Analytical Instruments 2009 165 of 236 DLM 90488 01 Rev A ...
Page 167: ...GE Analytical Instruments 2009 167 of 236 DLM 90488 01 Rev A ...
Page 170: ...GE Analytical Instruments 2009 170 of 236 DLM 90488 01 Rev A ...
Page 194: ...GE Analytical Instruments 2009 194 of 236 DLM 90488 01 Rev A ...
Page 218: ...GE Analytical Instruments 2009 218 of 236 DLM 90488 01 Rev A ...
Page 235: ...Notes GE Analytical Instruments 2009 235 of 236 DLM 90488 01 Rev A 236 ...
Page 236: ...Notes GE Analytical Instruments 2009 236 of 236 DLM 90488 01 Rev A 236 ...