GE Analytical Instruments ©2009
147 of 236
DLM 90488-01 Rev. A
If you have performed a reagent flush in the last 8 hours, click
No
to skip the flush and
proceed to Step 6.
6. When the flush is complete, click
OK
to continue. Remove the protective cap from the
standards vials, and avoid touching the septa to protect against introducing foreign
particles.
At this point, the prompts on the Analyzer’s display screen will help guide you through the
calibration process.
7. Insert the Calibration Blank into the IOS System, and press
Next
.
8. When prompted, remove the Calibration Blank from the IOS System, insert the 1 ppm TOC
Calibration Standard (as KHP), and press
Next
.
9. When prompted, remove the 1 ppm TOC Calibration Standard (as KHP) from the IOS
System, insert the 5 ppm TOC Calibration Standard (as KHP), and press
Next
.
10. When prompted, remove the 5 ppm TOC Calibration Standard (as KHP) from the IOS
System, insert the 10 ppm TOC Calibration Standard (as KHP), and press
Next
.
11. When prompted, remove the 10 ppm TOC Calibration Standard (as KHP) from the IOS
System, insert the 25 ppm TOC Calibration Standard (as KHP), and press
Next
.
12. When prompted, remove the 25 ppm TOC Calibration Standard (as KHP) from the IOS
System, insert the 50 ppm TOC Calibration Standard (as KHP), and press
Next
.
13. After the 50 ppm TOC Calibration Standard (as KHP) has been analyzed, you will be
prompted to enter the IC standards. Remove the 50 ppm TOC Calibration Standard (as KHP)
from the IOS System, insert the 1 ppm IC Calibration Standard (as Na
2
CO
3
), and press
Next
.
14. When prompted, remove the 1 ppm IC Calibration Standard (as Na
2
CO
3
) from the IOS
System, insert the 5 ppm IC Calibration Standard (as Na
2
CO
3
), and press
Next
.
15. When prompted, remove the 5 ppm IC Calibration Standard from the IOS System, insert the
10 ppm IC Calibration Standard (as Na
2
CO
3
), and press
Next
.
16. When prompted, remove the 10 ppm IC Calibration Standard from the IOS System, insert
the 25 ppm IC Calibration Standard (as Na
2
CO
3
), and press
Next
.
17. When prompted, remove the 25 ppm IC Calibration Standard (as Na
2
CO
3
) from the IOS
System, insert the 50 ppm IC Calibration Standard (as Na
2
CO
3
), and press
Next
.
18. After the 50 ppm IC Calibration Standard (as Na
2
CO
3
) has been analyzed, the calibration
summary screen is displayed (see Figure 25 and Figure 26). Toggle between the values for
each channel by pressing the
IC Data
and
TC Data
buttons. The Analyzer indicates if the
calibration passed or failed based on the following criteria:
Summary of Contents for Sievers 900 Series
Page 10: ...GE Analytical Instruments 2009 10 of 236 DLM 90488 01 Rev A ...
Page 88: ...GE Analytical Instruments 2009 88 of 236 DLM 90488 01 Rev A ...
Page 122: ...GE Analytical Instruments 2009 122 of 236 DLM 90488 01 Rev A ...
Page 159: ...GE Analytical Instruments 2009 159 of 236 DLM 90488 01 Rev A ...
Page 161: ...GE Analytical Instruments 2009 161 of 236 DLM 90488 01 Rev A ...
Page 165: ...GE Analytical Instruments 2009 165 of 236 DLM 90488 01 Rev A ...
Page 167: ...GE Analytical Instruments 2009 167 of 236 DLM 90488 01 Rev A ...
Page 170: ...GE Analytical Instruments 2009 170 of 236 DLM 90488 01 Rev A ...
Page 194: ...GE Analytical Instruments 2009 194 of 236 DLM 90488 01 Rev A ...
Page 218: ...GE Analytical Instruments 2009 218 of 236 DLM 90488 01 Rev A ...
Page 235: ...Notes GE Analytical Instruments 2009 235 of 236 DLM 90488 01 Rev A 236 ...
Page 236: ...Notes GE Analytical Instruments 2009 236 of 236 DLM 90488 01 Rev A 236 ...