GE Analytical Instruments ©2009
155 of 236
DLM 90488-01 Rev. A
Figure 28: The Autoreagent Calibration Summary Screen
Performing an Autoreagent Verification
Use the Autoreagent verification to confirm the Analyzer's current Autoreagent calibration is
accurate. GE Analytical Instruments recommends performing the Autoreagents verification after
replacement of pump tubing and after Autoreagent calibration.
The sample flow rate calibration (see page 141) should always be performed after replacement of
the pump tubing and before the Autoreagents verification.
To perform the Autoreagent calibration verification, follow this procedure:
1. If the Analyzer is taking measurements, press the
Stop Analysis
button.
2. If you have an ICR unit, make sure the lever is in the
Bypass
position.
3. Press the
Menu
button, select the
Maintenance
tab, and then press the
Calibrate
button.
4. Press the
Autoreagent Ver
button.
5. Remove the protective cap from the standards vials, and avoid touching the septa to
protect against introducing foreign particles.
At this point, the prompts on the Analyzer's display screen will help guide you through the
Autoreagent calibration verification process.
6. Insert the Reagent Water into the IOS System, and press
Next
.
Summary of Contents for Sievers 900 Series
Page 10: ...GE Analytical Instruments 2009 10 of 236 DLM 90488 01 Rev A ...
Page 88: ...GE Analytical Instruments 2009 88 of 236 DLM 90488 01 Rev A ...
Page 122: ...GE Analytical Instruments 2009 122 of 236 DLM 90488 01 Rev A ...
Page 159: ...GE Analytical Instruments 2009 159 of 236 DLM 90488 01 Rev A ...
Page 161: ...GE Analytical Instruments 2009 161 of 236 DLM 90488 01 Rev A ...
Page 165: ...GE Analytical Instruments 2009 165 of 236 DLM 90488 01 Rev A ...
Page 167: ...GE Analytical Instruments 2009 167 of 236 DLM 90488 01 Rev A ...
Page 170: ...GE Analytical Instruments 2009 170 of 236 DLM 90488 01 Rev A ...
Page 194: ...GE Analytical Instruments 2009 194 of 236 DLM 90488 01 Rev A ...
Page 218: ...GE Analytical Instruments 2009 218 of 236 DLM 90488 01 Rev A ...
Page 235: ...Notes GE Analytical Instruments 2009 235 of 236 DLM 90488 01 Rev A 236 ...
Page 236: ...Notes GE Analytical Instruments 2009 236 of 236 DLM 90488 01 Rev A 236 ...