17. Press the
Login
button and enter the new User ID and Password you just created.
18. Select the
DataGuard
tab.
19. Press the
Edit User
button.
20. Use the arrow buttons to highlight the default administrator account and press
OK
.
21. Press the
User Status
button and select
Inactive
.
22. Press the
Menu
button to return to the
Menu
screen.
Adding User IDs
After DataGuard has been enabled, you can add a User ID by following these steps:
1. Select the
DataGuard
tab.
2. Press the
Add User
button.
3. Specify a new User ID and press the
Enter
button. (If you enter an existing User ID, an error
message will display.)
4. Specify the temporary Password for the new User ID and press
Enter
. Unless you change
the expiration status (see “Editing User Information” on page 128), the user will be required
to change the Password when logging in for the first time.
5. Verify the Password and press the
Enter
button.
6. The
Edit User
screen displays automatically. Press the
Menu
button to accept the default
user settings, or edit any of the user settings. For additional information, see Step 4 in
“Editing User Information” below.
Editing User Information
Once you have created a User ID, you can edit its account settings as follows. Note that when you
are adding a new User ID, you do not need to follow Steps 1 through 3, as the
Edit User
screen
automatically displays whenever a new User ID is created.
1. Select the
DataGuard
tab.
Note:
Once you inactivate the default administrator account, it cannot be used to access the
Analyzer. If you forget the login information for the new administrator account that
you have created, contact GE Analytical Instruments for assistance.
Summary of Contents for Sievers 900 Series
Page 10: ...GE Analytical Instruments 2009 10 of 236 DLM 90488 01 Rev A ...
Page 88: ...GE Analytical Instruments 2009 88 of 236 DLM 90488 01 Rev A ...
Page 122: ...GE Analytical Instruments 2009 122 of 236 DLM 90488 01 Rev A ...
Page 159: ...GE Analytical Instruments 2009 159 of 236 DLM 90488 01 Rev A ...
Page 161: ...GE Analytical Instruments 2009 161 of 236 DLM 90488 01 Rev A ...
Page 165: ...GE Analytical Instruments 2009 165 of 236 DLM 90488 01 Rev A ...
Page 167: ...GE Analytical Instruments 2009 167 of 236 DLM 90488 01 Rev A ...
Page 170: ...GE Analytical Instruments 2009 170 of 236 DLM 90488 01 Rev A ...
Page 194: ...GE Analytical Instruments 2009 194 of 236 DLM 90488 01 Rev A ...
Page 218: ...GE Analytical Instruments 2009 218 of 236 DLM 90488 01 Rev A ...
Page 235: ...Notes GE Analytical Instruments 2009 235 of 236 DLM 90488 01 Rev A 236 ...
Page 236: ...Notes GE Analytical Instruments 2009 236 of 236 DLM 90488 01 Rev A 236 ...