GE Analytical Instruments ©2009
142 of 236
DLM 90488-01 Rev. A
9. The Analyzer will pump water into the 10-mL graduated cylinder for exactly 10 min. Then
the Analyzer will stop pumping, and it will ask you if the collection of the water was
satisfactory (for example, no spilling occurred no substantial air bubbles were observed).
If you want to repeat the water collection, dry the graduated cylinder and press
Repeat
. If
the water collection was satisfactory, press
Next
to continue.
10. Measure the volume of water in the graduated cylinder to the nearest 0.1 mL. Using the
keypad on the display, enter the volume and press
Enter
. The volume should be between
3.0 and 7.0 mL; if it is not, follow the on-screen prompts to resolve the problem.
11. The Analyzer will automatically adjust the sample flow rate, and it will request you to verify
that the adjustment is accurate. To do so, first drain and dry the 10-mL graduated
cylinder. After you replace the waste line into the graduated cylinder, press
Next
to
continue.
12. The Analyzer will pump water into the 10-mL graduated cylinder for exactly 10 min. Then
the Analyzer will stop pumping, and it will ask you if the collection of the water was
satisfactory. If you want to repeat the water collection, dry the graduated cylinder and
press
Repeat
. If the water collection was satisfactory, press
Next
to continue.
13. Measure the volume of water in the graduated cylinder to the nearest 0.1 mL. Enter the
volume using the keypad on the display, then click
Enter
.
14. The sample flow rate calibration summary screen is then displayed (see Figure 22). The
Analyzer indicates if the sample flow rate calibration passed or failed based on the
following criterion:
• the adjusted sample flow rate must be between 0.48 and 0.52 mL/min
If this condition is satisfied, press
Apply
to accept the sample flow rate calibration and
continue.
If this condition is not satisfied, press the
Cancel
button to reject the sample flow rate
calibration. You may need to perform the sample flow rate calibration procedure again.
However, first check that the sample tubing has been installed correctly. Also, look for any
leaks inside the Analyzer and correct them. Consult the chapter called "Troubleshooting"
(starting on page page 195) to determine if there is another problem with the Analyzer.
15. Disconnect the 12" waste line and restore the original waste line.
To remove the 12" waste line, use a 7/16" open-end wrench to loosen the fitting from the
Waste
port on the side of the Analyzer; store the 12" waste line in a secure, clean place.
Then, reattach the stainless steel tubing by removing the rubber plug and securing the 7/
16" nut to the
Waste
port.
Summary of Contents for Sievers 900 Series
Page 10: ...GE Analytical Instruments 2009 10 of 236 DLM 90488 01 Rev A ...
Page 88: ...GE Analytical Instruments 2009 88 of 236 DLM 90488 01 Rev A ...
Page 122: ...GE Analytical Instruments 2009 122 of 236 DLM 90488 01 Rev A ...
Page 159: ...GE Analytical Instruments 2009 159 of 236 DLM 90488 01 Rev A ...
Page 161: ...GE Analytical Instruments 2009 161 of 236 DLM 90488 01 Rev A ...
Page 165: ...GE Analytical Instruments 2009 165 of 236 DLM 90488 01 Rev A ...
Page 167: ...GE Analytical Instruments 2009 167 of 236 DLM 90488 01 Rev A ...
Page 170: ...GE Analytical Instruments 2009 170 of 236 DLM 90488 01 Rev A ...
Page 194: ...GE Analytical Instruments 2009 194 of 236 DLM 90488 01 Rev A ...
Page 218: ...GE Analytical Instruments 2009 218 of 236 DLM 90488 01 Rev A ...
Page 235: ...Notes GE Analytical Instruments 2009 235 of 236 DLM 90488 01 Rev A 236 ...
Page 236: ...Notes GE Analytical Instruments 2009 236 of 236 DLM 90488 01 Rev A 236 ...