1. If the Analyzer is taking measurements, press the
Stop Analysis
button.
2. Make sure the Analyzer is connected to a water supply.
3. Press the
Menu
button and select the
Maintenance
tab.
4. Press the
Consumables
button.
5. Press the
Flush...
button.
6. Press either the
Acid
,
Oxid
, or
Both
button, depending on which syringe(s) you want to
flush.
7. Make sure a water source is connected to the Analyzer and press the
Flush
button.
8. Wait for the flush to complete. The
Flush
button will change colors to show that the
process is underway and the Analyzer will not respond to additional commands. Do not
disrupt power to the Analyzer while the flush is in progress.
9. When the flush is complete, a message will display. Press
OK
to continue.
10. If you used a vial for the flush, remove it from the IOS System after the flush has finished.
Cleaning the Analyzer
You can clean the external housing of the Analyzer with a damp cloth using water or non-abrasive
cleaners. Turn off power to the Analyzer and disconnect it from main power prior to cleaning. Do
not spray liquids directly on the Analyzer. Wipe dry with a clean, soft cloth.
The touch screen may be cleaned with a soft cloth or cotton pad. Methanol, or Isopropyl Alcohol
may be used, but insure that all solvent residue is removed. Water may cause damage or
discoloration of the polarizer. Clean any condensation or moisture from any source immediately.
Before cleaning the screen, you can temporarily deactivate touch screen sensitivity for 30 seconds
by selecting the following option:
1. Press the
Menu
button and select the
Maintenance
tab.
2. Press the
Advanced
button, press the
Advanced Setup
button, and then press the
Clean
Screen
button.
Summary of Contents for Sievers 900 Series
Page 10: ...GE Analytical Instruments 2009 10 of 236 DLM 90488 01 Rev A ...
Page 88: ...GE Analytical Instruments 2009 88 of 236 DLM 90488 01 Rev A ...
Page 122: ...GE Analytical Instruments 2009 122 of 236 DLM 90488 01 Rev A ...
Page 159: ...GE Analytical Instruments 2009 159 of 236 DLM 90488 01 Rev A ...
Page 161: ...GE Analytical Instruments 2009 161 of 236 DLM 90488 01 Rev A ...
Page 165: ...GE Analytical Instruments 2009 165 of 236 DLM 90488 01 Rev A ...
Page 167: ...GE Analytical Instruments 2009 167 of 236 DLM 90488 01 Rev A ...
Page 170: ...GE Analytical Instruments 2009 170 of 236 DLM 90488 01 Rev A ...
Page 194: ...GE Analytical Instruments 2009 194 of 236 DLM 90488 01 Rev A ...
Page 218: ...GE Analytical Instruments 2009 218 of 236 DLM 90488 01 Rev A ...
Page 235: ...Notes GE Analytical Instruments 2009 235 of 236 DLM 90488 01 Rev A 236 ...
Page 236: ...Notes GE Analytical Instruments 2009 236 of 236 DLM 90488 01 Rev A 236 ...