GE Analytical Instruments ©2009
166 of 236
DLM 90488-01 Rev. A
System Suitability Worksheet
Company Name ___________________________
Date ________________________
Analyst Name ____________________________
Firmware Version _____________
Analyzer Serial Number ____________________
Average TOC of Reagent Water (Rw)
__________________
(Line 1)
Average TOC of 500 ppb Sucrose Standard (Rs)
__________________
(Line 2)
Average TOC of 500 ppb Benzoquinone Standard (Rss) __________________
(Line 3)
Response Efficiency
__________________
(Line 4)
Response Limit
__________________
(Line 5)
Response Limit = Rs - Rw
Acceptance criteria for USP System Suitability is response efficiency between 85% and 115%
Pass
Fail
Indicate whether the response efficiency results were applied or canceled:
Apply
Cancel
Response Efficiency = (Rss - Rw)
(Rs-Rw)
----------------------------
100
×
Summary of Contents for Sievers 900 Series
Page 10: ...GE Analytical Instruments 2009 10 of 236 DLM 90488 01 Rev A ...
Page 88: ...GE Analytical Instruments 2009 88 of 236 DLM 90488 01 Rev A ...
Page 122: ...GE Analytical Instruments 2009 122 of 236 DLM 90488 01 Rev A ...
Page 159: ...GE Analytical Instruments 2009 159 of 236 DLM 90488 01 Rev A ...
Page 161: ...GE Analytical Instruments 2009 161 of 236 DLM 90488 01 Rev A ...
Page 165: ...GE Analytical Instruments 2009 165 of 236 DLM 90488 01 Rev A ...
Page 167: ...GE Analytical Instruments 2009 167 of 236 DLM 90488 01 Rev A ...
Page 170: ...GE Analytical Instruments 2009 170 of 236 DLM 90488 01 Rev A ...
Page 194: ...GE Analytical Instruments 2009 194 of 236 DLM 90488 01 Rev A ...
Page 218: ...GE Analytical Instruments 2009 218 of 236 DLM 90488 01 Rev A ...
Page 235: ...Notes GE Analytical Instruments 2009 235 of 236 DLM 90488 01 Rev A 236 ...
Page 236: ...Notes GE Analytical Instruments 2009 236 of 236 DLM 90488 01 Rev A 236 ...