Additional System Components
Microprocessor Controller and Electronics
All operations of the Analyzer are controlled by a 32-bit microprocessor. Six proprietary electronic
board assemblies monitor and control Analyzer functions:
•
Main board
including the 32-bit microprocessor, 1 MB of program memory, 1 MB of data
memory, battery-backed nonvolatile memory for operator settings, digital I/O, QVGA color
graphics controller, stepper motor controllers, printer output, RS-232 (serial) port, and
interface to USB storage devices
•
Analog-to-digital conversion board
with integrated signal conditioning circuits for two
conductivity and four temperature measurements
•
Color LCD QVGA display
with touch panel
•
Passive interconnect board
•
ID board
with nonvolatile memory for system specific coefficients
•
I/O board
with interfaces to external devices, via binary output, two alarms, and 4-20 mA
output
Data Outputs
The Analyzer has serial (RS-232), USB, and printer ports, allowing flexibility for exporting and
printing data. Real-time and historical data can be transferred from the Analyzer via the serial port
to a computer using communication software such as HyperTerminal (for details, see “Using
HyperTerminal” on page 228). Historical data can be transferred from the Analyzer via the USB port
to a USB flash storage device (provided) and then transferred to any computer that supports USB.
The Sievers 900 On-line TOC Analyzer also has one analog output (4-20 mA) which can be
customized to track specific data values, two alarms, binary output for use with a sample
sequencer, and binary input to facilitate remote start/stop functions.
Note:
The Analyzer cannot be directly connected to a computer via the USB port. Rather, the
Analyzer’s USB port can only be connected to a USB storage device, such as a USB flash
memory drive.
Summary of Contents for Sievers 900 Series
Page 10: ...GE Analytical Instruments 2009 10 of 236 DLM 90488 01 Rev A ...
Page 88: ...GE Analytical Instruments 2009 88 of 236 DLM 90488 01 Rev A ...
Page 122: ...GE Analytical Instruments 2009 122 of 236 DLM 90488 01 Rev A ...
Page 159: ...GE Analytical Instruments 2009 159 of 236 DLM 90488 01 Rev A ...
Page 161: ...GE Analytical Instruments 2009 161 of 236 DLM 90488 01 Rev A ...
Page 165: ...GE Analytical Instruments 2009 165 of 236 DLM 90488 01 Rev A ...
Page 167: ...GE Analytical Instruments 2009 167 of 236 DLM 90488 01 Rev A ...
Page 170: ...GE Analytical Instruments 2009 170 of 236 DLM 90488 01 Rev A ...
Page 194: ...GE Analytical Instruments 2009 194 of 236 DLM 90488 01 Rev A ...
Page 218: ...GE Analytical Instruments 2009 218 of 236 DLM 90488 01 Rev A ...
Page 235: ...Notes GE Analytical Instruments 2009 235 of 236 DLM 90488 01 Rev A 236 ...
Page 236: ...Notes GE Analytical Instruments 2009 236 of 236 DLM 90488 01 Rev A 236 ...