Calibration pass/fail based upon the following criteria:
•
The R
2
is
≥
0.990.
•
The % Diff value for each IC standard is
±10% or less.
Calibration Results:
Passed
Failed
Calibration Action:
Applied
Canceled
% Diff = Measured Concentration - Expected Standard Concentration
Expected Standard Concentration
-----------------------------------------------------------------------------------------------------------------------------------------------------------------
100%
×