![GE Sievers 900 Series Operation And Maintenance Manual Download Page 177](http://html.mh-extra.com/html/ge/sievers-900-series/sievers-900-series_operation-and-maintenance-manual_128448177.webp)
GE Analytical Instruments ©2009
177 of 236
DLM 90488-01 Rev. A
To replace the UV lamp:
1. If the Analyzer is taking measurements, press the
Stop Analysis
button.
2. Turn off the Analyzer by using the main power switch.
3. Open the Analyzer front panel by turning the two latches and swinging the panel door to
the left.
4. Locate the UV lamp enclosure at the right of the Analyzer. The UV enclosure has a brass
nut at the bottom of the enclosure and a gray wire with a black connector extending from
the nut.
Warning
The UV lamp contains mercury and may be considered hazardous material in your
local area. Dispose of the UV lamp in accordance with federal, state, or local
government regulations.
Warning
Installation of the UV lamp requires access to the inside of the
Analyzer. To avoid potentially dangerous shock, disconnect the power
cord before opening the Analyzer’s front panel.
Warning
Should the UV lamp become broken or damaged, it should be handled in
accordance with your organization's toxic waste handling procedure and disposed
of in accordance with federal, state, or local government regulations.
Note:
A cotton glove to be worn during the installation is included with the replacement UV
lamp to avoid leaving fingerprints on the quartz window of the lamp. Fingerprints
absorb UV radiation and decrease the performance of the oxidation reactor. If
necessary, use methanol to remove any fingerprints before installing the new lamp.
Summary of Contents for Sievers 900 Series
Page 10: ...GE Analytical Instruments 2009 10 of 236 DLM 90488 01 Rev A ...
Page 88: ...GE Analytical Instruments 2009 88 of 236 DLM 90488 01 Rev A ...
Page 122: ...GE Analytical Instruments 2009 122 of 236 DLM 90488 01 Rev A ...
Page 159: ...GE Analytical Instruments 2009 159 of 236 DLM 90488 01 Rev A ...
Page 161: ...GE Analytical Instruments 2009 161 of 236 DLM 90488 01 Rev A ...
Page 165: ...GE Analytical Instruments 2009 165 of 236 DLM 90488 01 Rev A ...
Page 167: ...GE Analytical Instruments 2009 167 of 236 DLM 90488 01 Rev A ...
Page 170: ...GE Analytical Instruments 2009 170 of 236 DLM 90488 01 Rev A ...
Page 194: ...GE Analytical Instruments 2009 194 of 236 DLM 90488 01 Rev A ...
Page 218: ...GE Analytical Instruments 2009 218 of 236 DLM 90488 01 Rev A ...
Page 235: ...Notes GE Analytical Instruments 2009 235 of 236 DLM 90488 01 Rev A 236 ...
Page 236: ...Notes GE Analytical Instruments 2009 236 of 236 DLM 90488 01 Rev A 236 ...