GE Analytical Instruments ©2009
221 of 236
DLM 90488-01 Rev. A
3. Press the
Turbo
button and then press
On
to enable Turbo sampling mode. If you have
previously set
Turbo
to
On
, you can press the
Off
button to return the Analyzer to normal
sampling mode.
Instruments that have been calibrated for Turbo operation also can be used in normal,
non-Turbo mode without affecting measurements. However, instruments that only have
been calibrated for normal operation will not make accurate Turbo measurements until
they are calibrated for Turbo mode, as described below.
Reagent Flow Rates
Operating in Turbo mode requires a change to the reagent flow rates. The Autoreagent function
cannot be used in Turbo mode.
The reagent flow rates for Turbo mode should be set as follows:
Acid = 2.0 µL/min
Oxidizer = 2.0 µL/min
Performing a Turbo Calibration
Calibration for Analyzers running in Turbo mode is different from basic calibration. Because of the
increased volume of standards required for Turbo calibration, the IOS System is not used, and
instead standards are sampled directly from the 250 mL containers provided by GE Analytical
Instruments. In addition, only a single-point calibration is performed. The Analyzer must be in Turbo
mode to perform the calibration.
To calibrate an Analyzer in Turbo mode, follow these steps:
1. If the Analyzer is taking measurements, press the
Stop Analysis
button.
2. If you have an ICR unit, place the lever into the
Bypass
position.
3. Press the
Menu
button and then press the
Consumables
button.
4. Press the
Flush
button and then
Flush Both
button.
Note:
Turbo sampling is available only in On-Line Mode. See “Setting the Analyzer Mode” on
page 93 for related information.
Summary of Contents for Sievers 900 Series
Page 10: ...GE Analytical Instruments 2009 10 of 236 DLM 90488 01 Rev A ...
Page 88: ...GE Analytical Instruments 2009 88 of 236 DLM 90488 01 Rev A ...
Page 122: ...GE Analytical Instruments 2009 122 of 236 DLM 90488 01 Rev A ...
Page 159: ...GE Analytical Instruments 2009 159 of 236 DLM 90488 01 Rev A ...
Page 161: ...GE Analytical Instruments 2009 161 of 236 DLM 90488 01 Rev A ...
Page 165: ...GE Analytical Instruments 2009 165 of 236 DLM 90488 01 Rev A ...
Page 167: ...GE Analytical Instruments 2009 167 of 236 DLM 90488 01 Rev A ...
Page 170: ...GE Analytical Instruments 2009 170 of 236 DLM 90488 01 Rev A ...
Page 194: ...GE Analytical Instruments 2009 194 of 236 DLM 90488 01 Rev A ...
Page 218: ...GE Analytical Instruments 2009 218 of 236 DLM 90488 01 Rev A ...
Page 235: ...Notes GE Analytical Instruments 2009 235 of 236 DLM 90488 01 Rev A 236 ...
Page 236: ...Notes GE Analytical Instruments 2009 236 of 236 DLM 90488 01 Rev A 236 ...