Autoreagent Calibration Worksheet
Company Name ___________________________
Date ________________________
Analyst Name ____________________________
Firmware Version _____________
Analyzer Serial Number ____________________
Autoreagent calibration pass/fail based upon the following criteria:
•
The median peak time of the 50 ppm TOC Standard is from 20 to 100 seconds less than the
median peak time of the 25 ppm TOC Standard.
•
The RSDs of the valid expected IC and TC concentration measurements are 3% or less.
•
The RSDs of the IC and TC preliminary measurements are 5% or less
Autoreagent Calibration Results:
Passed
Failed
Autoreagent Calibration Action:
Applied
Canceled
Standard
Peak Time (sec)
25 ppm TOC
50 ppm TOC
25 ppm IC Standard
Expected Conc.
(ppm)
RSD (%)
Prelim. Meas.
(ppm)
RSD (%)
IC
TC
Summary of Contents for Sievers 900 Series
Page 10: ...GE Analytical Instruments 2009 10 of 236 DLM 90488 01 Rev A ...
Page 88: ...GE Analytical Instruments 2009 88 of 236 DLM 90488 01 Rev A ...
Page 122: ...GE Analytical Instruments 2009 122 of 236 DLM 90488 01 Rev A ...
Page 159: ...GE Analytical Instruments 2009 159 of 236 DLM 90488 01 Rev A ...
Page 161: ...GE Analytical Instruments 2009 161 of 236 DLM 90488 01 Rev A ...
Page 165: ...GE Analytical Instruments 2009 165 of 236 DLM 90488 01 Rev A ...
Page 167: ...GE Analytical Instruments 2009 167 of 236 DLM 90488 01 Rev A ...
Page 170: ...GE Analytical Instruments 2009 170 of 236 DLM 90488 01 Rev A ...
Page 194: ...GE Analytical Instruments 2009 194 of 236 DLM 90488 01 Rev A ...
Page 218: ...GE Analytical Instruments 2009 218 of 236 DLM 90488 01 Rev A ...
Page 235: ...Notes GE Analytical Instruments 2009 235 of 236 DLM 90488 01 Rev A 236 ...
Page 236: ...Notes GE Analytical Instruments 2009 236 of 236 DLM 90488 01 Rev A 236 ...