6. Wait for the TOC Autozero to complete before proceeding with the calibration.
Perform a Sample Flow Rate Calibration
Prior to calibrating the Analyzer, replace the pump tubing as part of the annual maintenance on the
Analyzer (see “Replacing the Sample Pump Tubing” on page 179). Then, perform a sample flow rate
calibration, as described in this section.
Before proceeding, make sure you have saved the current Analyzer system settings, as described in
“Export Current Constants” on page 139.
For the sample flow rate calibration, you will need the following items from the Analyzer's
accessories kit:
•
10-mL graduated cylinder
•
1/8" OD tubing with 1/8" Swagelok fitting for waste line, approximately 12" long
•
Plug for IOS waste line
Perform the sample flow rate calibration as follows:
1. If the Analyzer is taking measurements, press the
Stop Analysis
button.
2. Press the
Menu
button, select the
Maintenance
tab and press the
Calibrate
button.
3. Press the
Flow Rate Cal
button.
4. Ensure the Analyzer is connected to a source of water and then press
Next
to continue.
5. The Analyzer will then flush the acid and oxidizer out of the sample line.
6. When the flush is completed, the Analyzer will prompt guide you to attach the 12" waste
line (from the Analyzer’s accessories kit).
To install the 12" waste line, your must first disconnect the 7/16" nut that connects to the
port labeled
Waste
on the side of the Analyzer and gently pull the tubing aside; attach the
rubber plug to this tubing. Attach the 12" waste line by securing the fitting with a 7/16"
open-end wrench.
When the 12" waste line is installed, press
Next
to continue.
7. The Analyzer will fill the waste line with sample water.
8. When the waste line is filled, the Analyzer will instruct you to insert the end of the waste
line in the 10-mL graduated cylinder. Do so, and then press
Next
to continue.
Summary of Contents for Sievers 900 Series
Page 10: ...GE Analytical Instruments 2009 10 of 236 DLM 90488 01 Rev A ...
Page 88: ...GE Analytical Instruments 2009 88 of 236 DLM 90488 01 Rev A ...
Page 122: ...GE Analytical Instruments 2009 122 of 236 DLM 90488 01 Rev A ...
Page 159: ...GE Analytical Instruments 2009 159 of 236 DLM 90488 01 Rev A ...
Page 161: ...GE Analytical Instruments 2009 161 of 236 DLM 90488 01 Rev A ...
Page 165: ...GE Analytical Instruments 2009 165 of 236 DLM 90488 01 Rev A ...
Page 167: ...GE Analytical Instruments 2009 167 of 236 DLM 90488 01 Rev A ...
Page 170: ...GE Analytical Instruments 2009 170 of 236 DLM 90488 01 Rev A ...
Page 194: ...GE Analytical Instruments 2009 194 of 236 DLM 90488 01 Rev A ...
Page 218: ...GE Analytical Instruments 2009 218 of 236 DLM 90488 01 Rev A ...
Page 235: ...Notes GE Analytical Instruments 2009 235 of 236 DLM 90488 01 Rev A 236 ...
Page 236: ...Notes GE Analytical Instruments 2009 236 of 236 DLM 90488 01 Rev A 236 ...