GE Analytical Instruments ©2009
187 of 236
DLM 90488-01 Rev. A
22. Turn the Analyzer on with the main power switch.
23. Enter the installation date for the new resin bed. Press the
Menu
button, select the
Maintenance
tab, press the
Consumables
button, press the
New...
button, and then press
the
RBed
button (see “Setting the Installation or Expiration Date for New Consumables” on
page 190 for additional details, if necessary).
24. Before resuming normal operation, check for leaks.
After resuming normal operation it may take up to four hours for the resin bed to rinse
down; during this time you may experience higher than normal IC values. If high IC values
persist, then you should perform a Conductivity Autozero; follow the instructions in
“Conductivity Autozero” on page 211.
25. After about 12 hours of operation, check the interior of the Analyzer for leaks, and add
water to the DI water reservoir, if necessary.
Summary of Contents for Sievers 900 Series
Page 10: ...GE Analytical Instruments 2009 10 of 236 DLM 90488 01 Rev A ...
Page 88: ...GE Analytical Instruments 2009 88 of 236 DLM 90488 01 Rev A ...
Page 122: ...GE Analytical Instruments 2009 122 of 236 DLM 90488 01 Rev A ...
Page 159: ...GE Analytical Instruments 2009 159 of 236 DLM 90488 01 Rev A ...
Page 161: ...GE Analytical Instruments 2009 161 of 236 DLM 90488 01 Rev A ...
Page 165: ...GE Analytical Instruments 2009 165 of 236 DLM 90488 01 Rev A ...
Page 167: ...GE Analytical Instruments 2009 167 of 236 DLM 90488 01 Rev A ...
Page 170: ...GE Analytical Instruments 2009 170 of 236 DLM 90488 01 Rev A ...
Page 194: ...GE Analytical Instruments 2009 194 of 236 DLM 90488 01 Rev A ...
Page 218: ...GE Analytical Instruments 2009 218 of 236 DLM 90488 01 Rev A ...
Page 235: ...Notes GE Analytical Instruments 2009 235 of 236 DLM 90488 01 Rev A 236 ...
Page 236: ...Notes GE Analytical Instruments 2009 236 of 236 DLM 90488 01 Rev A 236 ...