Performing an Exam
2-58
LOGIQ V2/LOGIQ V1
–
User Guide
Direction 5610736-100
Rev. 9
Auto Trace
The value measured depends upon the Vol Flow Method preset.
The two selections available are: Peak (TAMAX) and Mean
(TAMEAN).
To auto trace TAMAX:
1. Press
Measure
key to enter into measure mode, if
necessary.
Adjust
Caliper Change
in the Primary Menu to
S/D Trace
.
In the Utility -> Measure - Advanced preset menu set
Trace
preset to
Auto
.
2. To position the caliper at the trace start point in the Doppler
spectrum, move the
Trackball
.
3. To fix the start point, press
Set
.
4. To position the vertical caliper at the end point, move the
Trackball
.
5. To complete the measurement, press
Set
. The system
automatically fixes both calipers and traces the maximum
value between the two points. The system displays this
value in the Results Window.