After the Exam is Over
3-60
LOGIQ V2/LOGIQ V1
–
User Guide
Direction 5610736-100
Rev. 9
Axial distance measurements
Description
Axial measurements are the distance measurements obtained
along the sound beam. See Figure 3-35 for more information.
Benefit
The accurate measurement of the size, depth and volume of a
structure is a critical factor in determining a proper diagnosis.
Most imaging systems use depth markers and/or electronic
calipers for this purpose.
Method
Axial distance should be measured in the near, mid and far fields
as well as in zoom. If necessary, different depths or fields of
view can be tested.
Procedure
1. Scan a test phantom with precisely-spaced vertical pin
targets. Adjust all scan controls, as necessary, for the best
image of the pin targets to typical depths for the probe being
used.
2. Press
Freeze
to stop image acquisition and perform a
standard distance measurement between the pins at
different points in the image. Record all images for
archiving.
3. Scan the vertical pins in zoom or at different depth/scale
factors.
4. Press
Freeze
to stop image acquisition; repeat the distance
measurements between pins and record the images for
archiving.
5. Document the measurements for reference and future
comparison.
Contact a Service Engineer if vertical measurements differ by
more than 5% of the actual distance.
Summary of Contents for LOGIQ V2
Page 8: ...i 6 LOGIQ V2 LOGIQ V1 User Guide Direction 5610736 100 Rev 9...
Page 92: ...Getting Started 1 80 LOGIQ V2 LOGIQ V1 User Guide Direction 5610736 100 Rev 9...
Page 242: ...After the Exam is Over 3 80 LOGIQ V2 LOGIQ V1 User Guide Direction 5610736 100 Rev 9...
Page 288: ...Safety 4 46 LOGIQ V2 LOGIQ V1 User Guide Direction 5610736 100 Rev 9...
Page 380: ...Index 4 LOGIQ V2 LOGIQ V1 User Guide Direction 5610736 100 Rev 9...
Page 381: ......