Preparing for a Biopsy
LOGIQ V2/LOGIQ V1
–
User Guide
5-29
Direction 5610736-100
Rev. 9
Displaying the Guidezone (continued)
NOTE:
You can display the biopsy guideline on the CFM image in
simultaneous mode. Select the Show Biopsy Mark on CFM
simultaneous Mode preset in the Utility -> System -> System
Image -> Biopsy Guide screen.
The biopsy guidezone represents a path of the needle. The dots
which make up the guidezones is the depth readout where:
•
Yellow represent 1 cm increments.
•
Red represents 5 cm increments.
The display should be carefully monitored during a biopsy for
any needle deviation from the center line or guidezone.
The Biopsy Guidezone adjusts along with image adjustments,
such as image inversion/rotations, zoom and depth changes.
The needle may vary from the center line or guidezone for
various reasons:
•
Needle barrel to needle clearance or strength.
• Bracket
manufacturing
tolerance.
•
Needle deflection due to tissue resistance.
•
Needle size chosen. Thinner needles may deflect more.
Table 5-13: Biopsy Guide Availability
Biopsy Guide
E8C-RS
L6
-1
2-RS
3Sc
-R
S
4C-RS
12
L
-R
S
Fixed Angle
E8C_TR5
Biopsy Depths (cm)
15.3
E8C_RU
Biopsy Angle (degree)
90
Multi-Angle
MBX1
Biopsy Depths (cm)
1.5
4.2
4.0
1.5
MBX2
Biopsy Depths (cm)
2.5
5.7
6.0
2.5
MBX3
Biopsy Depths (cm)
3.5
8.2
10.0
3.5
Summary of Contents for LOGIQ V2
Page 8: ...i 6 LOGIQ V2 LOGIQ V1 User Guide Direction 5610736 100 Rev 9...
Page 92: ...Getting Started 1 80 LOGIQ V2 LOGIQ V1 User Guide Direction 5610736 100 Rev 9...
Page 242: ...After the Exam is Over 3 80 LOGIQ V2 LOGIQ V1 User Guide Direction 5610736 100 Rev 9...
Page 288: ...Safety 4 46 LOGIQ V2 LOGIQ V1 User Guide Direction 5610736 100 Rev 9...
Page 380: ...Index 4 LOGIQ V2 LOGIQ V1 User Guide Direction 5610736 100 Rev 9...
Page 381: ......