Measurement and Analysis
LOGIQ V2/LOGIQ V1
–
User Guide
2-57
Direction 5610736-100
Rev. 9
TAMAX and TAMEAN
Manual Trace
The value measured depends upon the Vol Flow Method preset.
The two selections available are: Peak (TAMAX) and Mean
(TAMEAN).
To do a manual trace of TAMAX or TAMEAN:
1. Press
Measure
key to enter into measure mode, if
necessary.
Adjust
Caliper Change
in the Primary Menu to
S/D Trace
.
In the Utility -> Measure - Advanced preset menu set
Trace
preset to
Manual
.
2. To position the caliper at the trace start point, move the
Trackball
.
3. To fix the start point, press
Set
.
4. To trace the maximum values of the desired portion of the
spectrum, move the
Trackball
.
NOTE: To edit the trace line, move the
Trackball.
5. To complete the measurement, press
Set
. The system
displays the measurement values in the Results Window.
Summary of Contents for LOGIQ V2
Page 8: ...i 6 LOGIQ V2 LOGIQ V1 User Guide Direction 5610736 100 Rev 9...
Page 92: ...Getting Started 1 80 LOGIQ V2 LOGIQ V1 User Guide Direction 5610736 100 Rev 9...
Page 242: ...After the Exam is Over 3 80 LOGIQ V2 LOGIQ V1 User Guide Direction 5610736 100 Rev 9...
Page 288: ...Safety 4 46 LOGIQ V2 LOGIQ V1 User Guide Direction 5610736 100 Rev 9...
Page 380: ...Index 4 LOGIQ V2 LOGIQ V1 User Guide Direction 5610736 100 Rev 9...
Page 381: ......