Troubleshooting
LOGIQ 500 Basic Users Manual
2233658–100 Rev. 0
16–33
Operation Guide Message Description (cont’d)
Category
Problem/
Message
Possible
Cause
Possible
Corrective Action
Preset & Setup
Function
Data is saved.
The system completed
saving the data that was
requested.
NA
Input value by Rocker
Button.
Operator can use rocker
switch to input the value.
Use the Ellipse rocker
switch.
Input character strings.
Operator needs to input
some character strings.
Type in characters.
Select an initial data.
Operator needs to select
initial data.
Input proper data.
’TRACKBALL’ and ’SET’
are available.
Operator can use the
Trackball and Set key to
operate the function.
NA
Recommended value are
copyed.
The system completed
copying the data that
was requested.
NA
Table 16–5. Operation Guide Message Description (cont’d)
Warning Message Description
Category
Problem/
Message
Possible
Cause
Possible
Corrective Action
Temperature Safety
WARNING:
Temperature error.
System internal
temperature is at a
maximum, but not
caused by fan failure.
Reduce room
temperature.
WARNING
WARNING: Cooling Fan
Failure.
System fan has stopped
working properly.
Call service.
WARNING: Cooling Fan
Failure. Temperature
error.
System internal
temperature is at a
maximum, caused by the
fan not working properly.
Reduce room
temperature; call service.
WARNING: HV
Abnormal.
The High Voltage is
abnormal.
Call service.
WARNING: Probe
Temperature error.
The probe temperature
is at a maximum.
Call service.
WARNING: NOW
START THE POWER
OFF PROCESS.
The system shutdown
process has begun.
Call service.
Table 16–6. Warning Message Description