Probe Overview
LOGIQ 500 Basic Users Manual
2233658–100 Rev. 0
15–11
Special handling instructions
Using protective sheaths
Protective barriers may be required to minimize disease
transmission. Probe sheaths are available for use with all
clinical situations where infection is a concern. Use of legally
marketed, sterile probe sheaths is strongly recommended for
intra-cavitary and intra-operative procedures. Use of legally
marketed, sterile, pyrogen free probe sheaths is REQUIRED for
neurological intra-operative procedures.
Instructions.
Custom made sheaths are available for each
probe. Each probe sheath kit consists of a flexible sheath used
to cover the probe and cable and elastic bands used to secure
the sheath.
Sterile probe sheaths are supplied as part of biopsy kits for
those probes intended for use in biopsy procedures. In addition
to the sheath and elastic bands, there are associated
accessories for performing a biopsy procedure which are
included in the kit. Refer to the biopsy instructions for the
specific probes in the
Discussion section of this chapter for
further information.
Reordering.
To reorder sheaths, refer to
16–42.
Devices containing latex may cause severe allergic reaction in
latex sensitive individuals. Refer to FDA’s March 29, 1991
Medical Alert on latex products.
Do not use pre-lubricated condoms as a sheath. In some
cases, they may damage the probe. Lubricants in these
condoms may not be compatible with probe construction.
CAUTION
Biological
Hazard
CAUTION
CAUTION