Custom Display
LOGIQ 500 Basic Users Manual
2233658–100 Rev. 0
14–27
Page 8 of 18 (Imaging Parameter 8 – Probe Dependent 8) (cont’d)
PDI B Gain EC, UP, P, R
Choose the default B-Mode gain setting while in PDI Mode.
Value from 0 to 31 in 1 digit increments.
PDI B High Resolution EC, UP, P, R
Choose the initial default setting of the high resolution function in PDI Mode.
On or Off.
PDI B Frame Average EC, UP, P, R
Choose the level of frame averaging for B-Mode while in PDI.
Off
No B-Mode frame averaging.
Low
Low frame averaging.
Mid
Medium frame averaging.
High
High frame averaging.
PDI B Wall Filter EC, UP, P, R
Choose the default level of wall filtering for B-Mode while in PDI.
Low, M1, M2, M3, M4 and High.
PDI M Gain (Delta from B) EC, UP, P, R
Choose the default level for M-Mode gain while in PDI.
Level from –31 to 31 in 1 digit increments.
PDI M Wall Filter EC, UP, P, R
Choose the default level for the M-Mode Wall Filter while in PDI Mode.
Low, M1, M2, M3, M4, or High.
PDI ACE EC, UP, P, R
Choose the default setting for Adaptive Color Enhancement, which reduces clutter noise or
color flash.
ON or OFF.
PDI Noise Blanker EC, UP, P, R
Choose the default setting for Noise Blanker, which reduces random color noise.
OFF, LOW, MID or HIGH.