B–Mode
LOGIQ 500 Basic Users Manual
2233658–100 Rev. 0
5–4
Reading the B-Mode Display (cont’d)
LMP=xx/xx/xx
MHz
B-Mode Display
Description, Format, Values
Patient Name
A maximum of 29 alphanumeric characters.
Input at Patient Entry Menu.
ID
Patient identification number. A maximum of 14 alphanumeric characters.
Input at Patient Entry Menu.
Date
Today’s date according to the system settings.
Can be preset to display as MM/DD/YR, DD/MM/YR, or YR/MM/DD.
Time
Displays the current time during normal operation. Can be preset to a 12 or 24
hour clock. Displays the frame acquisition time when in Cine Mode.
Hospital Name
A maximum of 29 alphanumeric characters. Input at Set Up/System Parameters,
page 1.
LMP or GA
In GYN Exam Category:
In OB Exam Category:
LMP=xx/xx/xx
GA (xxx)=##W#D
The xxx after the GA shows how gestational age was calculated:
LMP, EDD or OPE (operator entered) – These are input in the Patient Entry Menu.
Acoustic Output
Percentage
The percentage is preceeded by a “P” and then a number from 0–100.
Frequency of
Operation
(Imaging Freq)
Frequency of Operation (for B- and M-Modes). The image frequency value is
displayed in MHz. The Frequency of Operation value coincides with the
“B Image
Frequency” and “M Image Frequency” presets located on Set Up/Custom Display
pages 2 and 3, respectively. The image frequency can range between Normal,
Resolution, Penetration, Further Penetration, Harmonic 1 and Harmonic 2. The
table on
5–19 shows the correlation between the image frequency optimization
setting and the frequency of operation value. When Harmonic is active, the letter
“T” is shown before the frequency value.
Probe
Probe name or designation of the active probe.
Scan Orientation
GE or
GE or the marker is used for scan orientation. This should coincide with the probe
orientation marking on the probe body. This marker can be turned off in Set
Up/Custom Display page 10.
Table 5–2. B-Mode Display Explanation