Custom Display
LOGIQ 500 Basic Users Manual
2233658–100 Rev. 0
14–30
Page 10 of 18 (Imaging Parameter 10)
Image Mode EC, R
Choose the default imaging mode for this preset.
B
B-Mode only
B/B
Dual B-Mode
B/M
B- and M-Modes
B/D
B- and Doppler Modes
CFM-B
B-Mode with Color Flow
CFM-B/B
B-Mode with Color Flow and B-Mode
CFM-B/M
B- and M-Modes with Color Flow
CFM-B/D
B- and Doppler Modes with Color Flow
Acoustic Power Knob Effect EC
Set how the Acoustic Output knob will affect the scan image.
Both B and D—Knob changes power for B/M/D/CFM/PDI all the time regardless of
scan mode.
Only B—Knob changes power for B or B/M mode only. D/CFM/PDI modes cannot be
adjusted by Acoustic power knob. If B or M-Modes are in combination with D/CFM/PDI
they also cannot be adjusted. It is fixed at setup preset value.
Independent B and D—Knob changes power for B/M-Mode while in B/M-Mode. Knob
changes power for D/CFM/PDI mode while in D/CFM/PDI mode.
No Effect—Acoustic output knob has no effect on changing power. Values fixed at set
up preset numbers.
Displayed TI EC
Choose the type of thermal index display.
TIS
Thermal Index – Soft Tissue
TIC
Thermal Index – Cranial Bone
TIB
Thermal Index – Bone
Image Gamma [%] EC, UP
Choose the level of gamma correction applied to the image during photography.
Choose from 50 to 200 in 5 digit increments.
Auto Focus Control EC
Choose to have the focus control follow depth changes as a percent of depth or have
the focus depth be a fixed value.
On
Focus is a percent of the current depth (Focus Pos % Depth preset) until
changed by focus position.
Off
Focus is at a fixed depth until changed by focus position control (Focus Depth
[mm] Preset).