Custom Display
LOGIQ 500 Basic Users Manual
2233658–100 Rev. 0
14–14
Page 1 of 18 (Imaging Parameter 1 – Probe Dependent 1) (cont’d)
Standoff Setting EC, UP, P
Choose a setting for the type of probe standoff, if used, for the 739L or L764 (LH)
probes. When any other probe is selected the preset is “Not Available”.
Off (No standoff), WP (Water path) or CPL (Couplant)
Needle Guide Type EC, UP, P
Choose the type of biopsy guidezone to be displayed for the guide attachment used.
For most probes, choose either Single Fixed Angle (SGL) or Multi Angle (MBX). For the
E721/MTZ, the probe guidezone choices are:
TV0
°
Reusable metal guide with a 0 degree offset angle.
TR5
°
Civco disposable guide with a 5 degree offset angle.
Needle Guide
Biopsy Zone
Guide Attachment
Type Setting
Soft Menu Choices
Multi Angle Attachment
MBX
MBX1, MBX2, MBX3, OFF
Fixed Angle Attachment
SGL
SGL, OFF
E721/MTZ Metal Guide
TV0
°
TV0
°
, OFF
E721/MTZ Disposable Guide
TR5
°
TR5
°
, OFF
ERB7 Disposable Guide
BX
1~8, OFF
ERB7 Needle Placement Guide
GBX
GBX, OFF
Failure to match the guidezone displayed to the guide
may cause the needle to track a path outside the
displayed zone.
It is extremely important that when using the adjustable
angle biopsy guides, the angle displayed on the screen
matches the angle set on the guide, otherwise the needle
will not follow the displayed guidezone which could result
in repeated biopsies or patient injury.
DANGER
Acoustic Power B/M [%] EC, UP, P, R
Set the initial percentage of acoustic power to be used when in B- or B/M-Modes.
Percentage from 0 to 100% in 10% increments.
Acoustic Power D/CFM/PDI [%] EC, UP, P, R
Set the initial percentage of acoustic power to be used when in Doppler, Color Flow or
Power Doppler imaging modes.
Percentage from 0 to 100% in 10% increments.