Probe Overview
LOGIQ 500 Basic Users Manual
2233658–100 Rev. 0
15–16
Disinfecting probes (cont’d)
Special Disinfecting Instructions for the E721(MTZ): To
properly disinfect the E721(MTZ) probe, the probe handle can
be reassembled loosely so that the entire probe with handle can
be immersed in the germicide solution. The adjustment screw
must be kept loose so that germicide can penetrate to all
surfaces. After immersing, rotate and shake the probe while it
is below the surface of the germicide to eliminate air pockets.
Allow the germicide to remain in contact with the fully immersed
probe, for high level disinfection, according to the germicide
manufacturer’s recommended time. To remove all germicide
residue, final rinsing should be done following the germicide
manufacturer’s instructions. Remove excess water by shaking
and allow to air dry.
CREUTZFIELD–JACOB DISEASE
Neurological use on patients with this disease must be avoided.
If a probe becomes contaminated, there is no adequate
disinfecting means.
.
CAUTION
Biological
Hazard