Preparing for a Biopsy
LOGIQ 500 Basic Users Manual
2233658–100 Rev. 0
15–33
Needle Guide Type Preset Selection
A preset is provided in Set Up/Custom Display page 1 to
choose single (SGL) or Multiple (MBX) guidezone angle
capabilities. Available selections for respective probes are
shown in the chart below.
Depth in cm at Center Channel
Fixed–
Angle
Multi–Angle
(MBX)
Probe
SGL
MBX1
MBX2
MBX3
C358
n/a
4.0
6.0
8.0
C364/CBF
8.0
4.0
6.0
8.0
C386
n/a
4.0
6.0
8.0
C551/CAE
7.0
4.0
5.5
7.0
C721
3.0
2.0
3.0
4.5
546L
n/a
4.0
5.5
7.0
L764/LH
2.0
2.0
4.0
7.0
739L
2.0
2.0
4.0
6.0
LA39
1.5
n/a
n/a
n/a
S316/UC
8.0
n/a
n/a
n/a
S317
8.0
4.0
6.0
8.0
S611
n/a
4.0
5.5
7.0
T739
2.0
2.0
4.0
6.0
Table 15–8. Biopsy Guide Attachment Selection
Failure to match the guidezone displayed to the guide may
cause the needle to track a path outside the zone.
It is extremely important that when using the adjustable angle
biopsy guides, the angle displayed on the screen matches the
angle set on the guide, otherwise the needle will not follow the
displayed guidezone which could result in repeated biopsies or
patient injury.
NOTE: Although the multi-angle guides are compatible with the
Civco Ultrapro and Ultrapro II, it is recommended the
multi-angle guides only be used with the Ultrapro II.
DANGER
.