3–6–2
LED List
The table below shows the LED list for LOGIQ
α
200.
TABLE 3–7
LED
LOCATION
NAME
DESCRIPTION
POSITION
NORMAL
ABNORMAL
MST ASSY
V5
+5V for digital
Edge of board
ON
OFF
WOTOUT
Watchdoc timer out, Reserved
Edge of board
–
–
D3~D10
Reserved
Edge of board
–
–
DSC ASSY
DS1
+5V for digital
Edge of board
ON
OFF
DS2
Mapping clock
Edge of board
Refer to
Note
NO
DS3
Mapping clock
Edge of board
Note
DS4
Mapping clock
Edge of board
DS5
DSP running
Edge of board
ON
OFF
HV ASSY
DS1
HV ASSY status (ON:normal)
Edge of board
ON
OFF
SMPS ASSY
GREEN
LED
SMPS ASSY status (ON:nor-
mal)
SMPS ASSY
ON
OFF
Note
Three LED (DS2~DS4) should be blinked when the system operated with the Probe.
3–6–3
DIP Switch
The table below shows the DIP Switch list for LOGIQ
α
200.
TABLE 3–8
DIP SWITCH Setting
LOCATION
NAME
Switch No.
DESCRIPTION
POSITION
MST ASSY
S1
1
Initiate the SRAM data to factory
Setup value
Edge of board
7
Select AAF Filter
(ON:Other countries, OFF:Japan)
2~ 6
Not Assigned
8
Select NTSC or PAL (ON:PAL, OFF:NTSC)
DSC ASSY
S1
1 ~8
Not Assigned
Edge of board
3–6–4
Reset Switch
Reset switch (S2) on edge of MST Assy used for resetting the system.