GE MEDICAL SYSTEMS
LOGIQ
α
200 SERVICE MANUAL
2138853
6–34
RENEWAL PARTS
REV 4
6–2–9
Keyboard Cover Set (FRU No. 203)
Time Required
11 Minutes
Tool Required
Screwdriver
Procedure
Refer to ILLUSTRATION 6–15.
1.
Turn OFF the system.
2.
Remove the Probe Holder (FRU No. 201). Refer to 6–2–7.
3.
Remove the Holder Bracket Assy (FRU No.202) Refer to 6–2–8.
4.
Unscrew two ground screws (1 and 2) on the front bottom of Keyboard .
5.
Unscrew three screws on the PCB that hold Ground Terminal wire and remove three clamps.
6.
Unscrew two screws (
) and remove two Ground Terminals.
7.
Disconnect the connector and cut the tie wraps off.
8.
Remove the KB Cover Top Assy with Keyboard Assy (FRU No. 204).
9.
Unscrew eight screws (3 – 10) and remove the KB Cover Top Assy.
10.
Unscrew nine screws (11 – 19) and remove the KB Cover Bottom.