GE MEDICAL SYSTEMS
LOGIQ
α
200 SERVICE MANUAL
2138853
6–52
RENEWAL PARTS
REV 5
6–2–21
Shield Panel (FRU No. 307)
Time Required
11 Minutes
Tool Required
Screwdriver
Procedure
CAUTION
Do NOT unplug the power cord before OFF the power switch.
Refer to ILLUSTRATION 6–27.
1.
Turn OFF the system & plug out
2.
Remove the Left Cover (FRU 404). Refer to 6–2–27.
3.
Remove the Right Cover (FRU 405). Refer to 6–2–28.
4.
Remove the Front Base Cover (FRU 406). Refer to 6–2–29.
5.
Remove the Front Cover (FRU 407). Refer to 6–2–30.
6.
Remove the Power S/W Assy (FRU 416). Refer to 6–2–38.
7.
Unscrew eleven screws (1 –11).
8.
Remove the Shield Panel (FRU 307).