GE MEDICAL SYSTEMS
LOGIQ
α
200 SERVICE MANUAL
2138853
6–40
RENEWAL PARTS
REV 4
6–2–13 Encoder Set (FRU No. 207)
Time Required
14 Minutes
Tool Required
Screwdriver, Hexagonal Wrench
Procedure
Refer to ILLUSTRATION 6–19
1.
Turn OFF the system.
2.
Pull the Keyboard Assy out with KB Cover Top Assy. Refer to 6–2–10
3.
Unscrew hexagonal screws (1) on each Encoder Knob and remove three knobs.
4.
Unscrew four screws on each Encoder Bracket and disconnect three connects (CN2 –CN4).
5.
Remove Encoder Set.
CN2
CN4
CN3
Encoder Bracket
1
ENCODER SET DISASSEMBLY
ILLUSTRATION 6–19