GE MEDICAL SYSTEMS
LOGIQ
α
200 SERVICE MANUAL
2138853
6–62
RENEWAL PARTS
REV 5
6–2–26
Pole Cover (FRU No. 403)
Time Required
20 Minutes
Tool Required
Screwdriver
Procedure
Refer to ILLUSTRATION 6–32.
1.
Turn OFF the system.
2.
Remove the Monitor Assy (FRU No. 100) from the Swing Arm Assy. Refer to 6–2–1.
3.
Remove the Swing Arm Assy (FRU 401). Refer to 6–2–24.
4.
Remove the Left Cover (FRU 404). Refer to 6–2–27.
5.
Remove the Right Cover(FRU 405). Refer to 6–2–28.
6.
Remove the Rear Cover (FRU 408). Refer to 6–2–31.
7.
Unscrew two screws (1and 2).
8.
Pull upward the Pole Cover.