GE MEDICAL SYSTEMS
LOGIQ
α
200 SERVICE MANUAL
2138853
6–108
RENEWAL PARTS
REV 5
6–4–2
Replacement Procedure (Continued)
(C) Replacing SMPS Assy
1.
Unscrew two hexagonal bolts (1 and 2). Two bolts are located at the front of SMPS Assy.
2.
Pull out the SMPS Assy forewards. Refer to ILLUSTRATION 6–58.
WARNING!
PERSONAL INJURY HAZARD! SMPS ASSY UNIT WEIGHTS APPROXIMATELY 3.5
KILOGRAMS (7.8 POUNDS). PAY ATTENTION NOT TO BE INJURED.
3.
Remove the SMPS Assy and place new one on the pallet.
4.
Slide and move the pallet to the position just under the console. Make sure that the SMPS Assy unit is placed
in the correct direction.
5.
Screw two hexagonal screws to assemble new SMPS Assy unit to the console.
ÇÇÇ
ÇÇÇ
ÇÇÇ
ÇÇÇ
ÇÇÇ
1
2
SMPS ASSY UNIT REMOVAL
ILLUSTRATION 6–58