INSTALLATION

2–14

REV 2

2–2–7

Transducer Connection

1.

Connect a transducer to the left side connector slot #1, transducer receptacle as follows:

a. Ensure that the transducer twist lock lever points towards the 12 o’clock position.

b. Insert the transducer connector on the receptacle guide pin until it touches the receptacle mating surface.

c. Twist the transducer twist lock lever to the 4 o’clock position to lock it in place.  Twist the lever to the 12

o’clock position to disconnect the transducer.

Note

It is not necessary to turn OFF power to connect or disconnect a transducer.

2.

Connect the main power cable to a hospital grade power receptacle with the proper rated voltage checked
during preinstallation.  Never use a three–to–two prong adapter; this defeats the safety ground.

2–2–8

Moving into Position

CAUTION

Do not lift the unit by the Keyboard.
Do not tilt the unit more than 5 degrees to avoid tipping it over.
To avoid injury by tipping over.  Set the monitor to the lowest position before moving.

CAUTION

Equipment Damage Possibility.  Lifting the console by holding covers may damage the
covers.  Do not lift the console by holding any covers.

In general, a single adult can move the LOGIQ

 

α

200.  (It is better to pull from the rear rather than push from the

front of the unit).  Before moving, store all loose parts in the unit.  Wrap transducers in soft cloth or foam to
prevent damage.

2–2–9

Adjusting System Clock

Set system clock for the LOGIQ

 

α

200to the local time. For procedure of adjusting the System clock, refer to

“Customizing Your System” in the Users manual.

Summary of Contents for LOGIQ ?200

Page 1: ... α ...

Page 2: ... damage MUST be reported to the carrier immediately upon discovery or in any event within 14 days after receipt and the contents and containers held for inspection by the carrier A transportation company will not pay a claim for damage if the inspection is not requested within this 14 day period GE Medical Systems Americas USA 1 800 437 1171 Canada 1 800 668 0732 Latin South Am 1 305 735 2304 GE M...

Page 3: ...R D Support Group Of SGMS FAX Number 82 342 420 420 REV 0 2184997 ...

Page 4: ......

Page 5: ...SAMSUNG GE MEDICAL SYSTEMS LTD SERVICE TEAM 64 3 Sangdaewon dong Chungwon ku Sungnam si Kyunggido KOREA ...

Page 6: ...ORTED DO NOT TEAR FOLD HERE AND SEAL DO NOT TEAR SERVICE RECORD SUBMITTED YES NO PLEASE BE SPECIFIC IN YOUR CORRECTIONS AND SUGGESTIONS FOLD FOLD IF ADDITIONAL PAGES ARE INCLUDED FOLD AND STAPLE TO LOWER PART OF THIS SHEET BEFORE MAILING SERVICE REP PHONE FAX COUNTRY CITY REV0 Stick along this line 2184996 ...

Page 7: ...ES À DES DANGERS ÉLECTRIQUES MÉCANIQUES OU AUTRES DIESES KUNDENDIENST HANDBUCH EXISTIERT NUR IN ENGLISCHER SPRACHE FALLS EIN FREMDER KUNDENDIENST EINE ANDERE SPRACHE BENÖTIGT IST ES AUFGABE DES KUNDEN FÜR EINE ENTSPRECHENDE ÜBERSETZUNG ZU SORGEN VERSUCHEN SIE NICHT DAS GERÄT ZU REPARIEREN BEVOR DIESES KUNDENDIENST HANDBUCH NICHT ZU RATE GEZOGEN UND VERSTANDEN WURDE WIRD DIESE WARNUNG NICHT BEACHTE...

Page 8: ...IENTE DEVIDO A CHOQUES ELÉTRICOS MECÂNICOS OU OUTROS IL PRESENTE MANUALE DI MANUTENZIONE È DISPONIBILE SOLTANTO IN INGLESE SE UN ADDETTO ALLA MANUTENZIONE ESTERNO ALLA GEMS RICHIEDE IL MANUALE IN UNA LINGUA DIVERSA IL CLIENTE È TENUTO A PROVVEDERE DIRETTAMENTE ALLA TRADUZIONE SI PROCEDA ALLA MANUTENZIONE DELL APPARECCHIATURA SOLO DOPO AVER CONSULTATO IL PRESENTE MANUALE ED AVERNE COMPRESO IL CONTE...

Page 9: ...PAGE REV PAGE REV PAGE REV Title Page 9 GE Logo Page 0 2184996 0 2184997 0 2184998 0 A 9 i to vi 5 Chapter 1 1 1 to 1 2 2 1 3 to 1 7 0 1 8 to 1 13 9 1 14 to 1 16 0 1 17 3 1 18 to 1 19 0 1 20 9 Chapter 2 2 1 3 2 2 to 2 3 0 2 4 2 2 5 0 2 6 2 2 7 to 2 8 0 2 9 5 2 10 0 2 11 5 2 12 to 2 13 0 2 14 to 2 15 2 2 16 0 Chapter 3 3 1 2 3 2 to 3 3 0 3 4 to 3 6 3 3 7 2 3 8 3 3 9 0 3 10 3 3 11 2 3 12 0 3 13 5 3 ...

Page 10: ......

Page 11: ...mance 1 18 1 3 2 Notice upon Installation of Product 1 18 1 3 3 General Notice 1 19 1 3 4 Countermeasures against EMC related Issues 1 19 1 3 5 Notice on Service 1 19 1 4 ADDRESS 1 20 CHAPTER 2 INSTALLATION 2 1 PREINSTALLATION 2 3 2 1 1 Introduction 2 3 2 1 2 Power Line Requirements 2 3 2 1 3 Physical Specifications 2 4 2 1 4 Recommended Ultrasound Room Layout 2 5 2 2 INSTALLATION 2 8 2 2 1 Introd...

Page 12: ...ONS 3 4 3 3 1 Power Supply 3 4 3 3 2 Facility Power Receptacle 3 4 3 4 STORAGE AND OPERATION REQUIREMENTS 3 4 3 5 OPTIONAL PERIPHERALS 3 5 3 5 1 Peripherals Accessories Connector Panel 3 5 3 5 2 List of Optional Peripherals 3 8 3 5 3 Power Consumption of Optional Peripherals 3 9 3 6 TEST POINT LED DIP SWITCH AND RESET SWITCH 3 10 3 6 1 Test Point List The table below shows The Test Point list for ...

Page 13: ...CIRCUIT BOARD DESCRIPTION 5 7 CHAPTER 6 RENEWAL PARTS 6 1 RENEWAL PARTS 6 3 6 2 DISASSEMBLY RE ASSEMBLY 6 23 6 2 1 Monitor Assy FRU No 100 6 24 6 2 2 Pot Set Assy FRU No 101 6 26 6 2 3 CRT Filter FRU No 102 6 27 6 2 4 Monitor Cover Set FRU No 103 6 28 6 2 5 Tilt Assy FRU No 104 6 29 6 2 6 Monitor Space Plate FRU No 105 6 30 6 2 7 Probe Holder FRU No 201 6 31 6 2 8 Holder Bracket Assy FRU No 202 6 ...

Page 14: ...6 68 6 2 31 Rear Cover FRU No 408 6 70 6 2 32 Top Cover FRU No 409 6 72 6 2 33 Printer Cover FRU No 410 6 74 6 2 34 Printer Bracket Assy FRU No 411 Back Bracket Assy FRU No 412 6 76 6 2 35 Rear Handle FRU No 413 6 78 6 2 36 Neck Frame FRU No 414 6 80 6 2 37 Rear Panel Assy FRU No 415 6 82 6 2 38 Power S W Assy FRU No 416 6 84 6 2 39 AC Fan Assy FRU No 417 6 86 6 2 40 Front Caster FRU No 418 6 88 6...

Page 15: ...CEDURE 7 3 7 2 1 Visual Inspection 7 3 7 2 2 Cleaning 7 4 7 2 3 Measurement 7 4 7 2 4 Battery replacement 7 4 7 2 5 Note 7 4 7 3 ELECTRICAL SAFETY TESTS 7 5 7 3 1 Outlet Test Wiring Arrangement 7 5 7 3 2 Grounding Continuity 7 6 7 3 3 Chassis Leakage Current Test 7 7 7 3 4 Probe Leakage Current Test 7 10 7 3 5 When There s Too Much Leakage Current 7 14 CHAPTER 8 INSTALLATION FOR OPTIONS 8 1 INTROD...

Page 16: ...GE MEDICAL SYSTEMS LOGIQ α200 SERVICE MANUAL 2138853 vi TABLE OF CONTENTS REV 5 This page is left blank intentionally ...

Page 17: ...E 1 1 SERVICE MANUAL CONTENTS 1 3 1 2 SAFETY 1 4 1 2 1 Warnings 1 4 1 2 2 Specifications 1 17 1 3 EMC Electromagnetic Compatibility 1 18 1 3 1 EMC Performance 1 18 1 3 2 Notice upon Installation of Product 1 18 1 3 3 General Notice 1 19 1 3 4 Countermeasures against EMC related Issues 1 19 1 3 5 Notice on Service 1 19 1 4 ADDRESS 1 20 ...

Page 18: ...GE MEDICAL SYSTEMS LOGIQ α200 SERVICE MANUAL 2138853 INTRODUCTION 1 2 REV 2 This page is left blank intentionally ...

Page 19: ... 3 System Configuration Contains system configuration and specifications 4 Chapter 4 Functional Checks Contains functional checks that must be performed as part of the installation or as required during servicing and periodic maintenance 5 Chapter 5 Diagrams Contains block diagrams and functional explanations of the LOGIQ α200 electronics 6 Chapter 6 Renewal Parts Contains a complete list of repla...

Page 20: ...tor when LOGIQ α200 is being operated 5 Probes are fragile please handle with care 6 Concerning Outside Markings refer to Illustration 1 1 1 2 1 3 1 4 1 5 1 6 1 7 1 8 and 1 9 7 For the cleaning disinfection and sterilization refer to Probe section in LOGIQ α200 User Manual and Caution Sheet supplied with each probe NOTICE This medical equipment is approved in terms of the prevention of radio wave ...

Page 21: ...GE MEDICAL SYSTEMS LOGIQ α200 SERVICE MANUAL 2138853 INTRODUCTION 1 5 REV 0 1 2 1 Warnings Continued ËËË OUTSIDE MARKINGS OF LOGIQ α200 FOR ALL UNITS ILLUSTRATION 1 1 ...

Page 22: ...ERVICE MANUAL 2138853 INTRODUCTION 1 6 REV 0 1 2 1 Warnings Continued OUTSIDE MARKINGS OFLOGIQ α200 FOR ALL UNITS ILLUSTRATION 1 2 Note For further details regarding the cautions above refer to 2 2 8 MOVING INTO POSITION in Chapter 2 ...

Page 23: ...GE MEDICAL SYSTEMS LOGIQ α200 SERVICE MANUAL 2138853 INTRODUCTION 1 7 REV 0 1 2 1 Warnings Continued ÍÍÍ ÍÍÍ OUTSIDE MARKINGS OFLOGIQ α200 FOR ALL UNITS ILLUSTRATION 1 3 ...

Page 24: ...lass1 Classe1 MANUFACTURED LOCATION GE MEDICAL SYSTEMS KOREA S N VOLT 220 240V AMP LONG TERM 2 1A KVA 0 5kVA PHASE 1 DESC LOGIQ 200 CISPR 11 EN 55011 CLASS A GROUP 1 CLASSE A GROUPE 1 OUTSIDE MARKINGS OFLOGIQ α200 FOR EUROPE ILLUSTRATION 1 4 Note For the symbols shown in the illustration above refer to latter pages in this chapter ...

Page 25: ...SS 1 MODEL 2205675 SERIAL MANUFACTURED VOLTS 120Vac 1 PHASE POWER 500VA FREQUENCY 60HZ DESC LOGIQ 200 For USA Canada Mexico MADE FOR GE MEDICAL SYSTEMS MILWAKEE WISCONSIN BY GE MEDICAL SYSTEMS KOREA CLASS 1 MODEL 2205676 SERIAL MANUFACTURED VOLTS 220Vac 240Vac 1 PHASE POWER 500VA FREQUENCY 50HZ DESC LOGIQ 200 For Brazil OUTSIDE MARKINGS OFLOGIQ α200 FOR USA CANADA MAXICO BRAZIL ILLUSTRATION 1 5 ...

Page 26: ...LTS 220Vac 240Vac 1 PHASE POWER 500VA FREQUENCY 50HZ DESC LOGIQ 200 For SEA China Hongkong India Australia New Zealand MADE FOR GE MEDICAL SYSTEMS MILWAKEE WISCONSIN BY GE MEDICAL SYSTEMS KOREA CLASS 1 MODEL 2205671 SERIAL MANUFACTURED VOLTS 120Vac 1 PHASE POWER 500VA FREQUENCY 60HZ DESC LOGIQ 200 For Taiwan Philippines OUTSIDE MARKINGS OFLOGIQ α200 FOR SEA CHINA HONGKONG INDIA AUSTRALLIA NEWZEALA...

Page 27: ... 1 Warnings Continued ÍÍÍÍ ÍÍÍÍ ÍÍÍÍ ÍÍÍÍ MADE FOR GE MEDICAL SYSTEMS MILWAKEE WISCONSIN BY GE MEDICAL SYSTEMS KOREA CLASS 1 MODEL 2205677 SERIAL MANUFACTURED VOLTS 220 240Vac 1 PHASE POWER 500VA FREQUENCY 60HZ DESC LOGIQ 200 OUTSIDE MARKINGS OFLOGIQ α200 FOR CHILE ILLUSTRATION 1 7 ...

Page 28: ...GE MEDICAL SYSTEMS LOGIQ α200 SERVICE MANUAL 2138853 INTRODUCTION 1 12 REV 9 1 2 1 Warnings Continued ÍÍÍ ÍÍÍ ÍÍÍ ÍÍÍ OUTSIDE MARKINGS OFLOGIQ α200 FOR KOREA ILLUSTRATION 1 8 ...

Page 29: ...DUCTION 1 13 REV 9 1 2 1 Warnings Continued CAUTION Do not use a Defibrillator simultaneously with the ECG as its excessive voltage will damage the signal input block of the ECG unit ÍÍÍÍ ÍÍÍÍ OUTSIDE MARKINGS OF LOGIQ α200 FOR UNITS WITH ECG ILLUSTRATION 1 9 ...

Page 30: ... 2138853 INTRODUCTION 1 14 REV 0 1 2 1 Warnings Continued NOTE Two labels are attached on rear of the SMPS assy box inside the rear cover and front of HV Board Assy inside the front base cover MARKINGS OF LOGIQ 500 INSIDE COVERS ILLUSTRATION 1 10 ...

Page 31: ...he box symbol IEC 878 02 03 indicates B Type equipment having a floating applied part Probe connectors Equipment Type CF heart in the box symbol IEC 878 02 05 indicate equipment having a floating applied part having a degree of protection suitable for direct cardiac contact ECG connector and surgical probes Device Listing Certification Labels Laboratory logo or labels denoting conformance with ind...

Page 32: ...ains OFF Indicates the power off position of the mains power switch Front of system Main power switch Mains ON Indicates the power on position of the mains power switch Front of system Main power switch Equipotentiality Indicates the terminal to be used for connecting equipotential conductors when interconnecting grounding with other equipment CAUTION This is only for FUNCTIONAL GROUNDING NOT PROT...

Page 33: ...ixed wiring of the installation in such a way that ACCESSIBLE METAL PARTS cannot become LIVE in the event of a failure of the BASIC INSULATION 2 Type BF EQUIPMENT TYPE B EQUIPMENT with an F TYPE APPLIED PART TYPE B EQUIPMENT EQUIPMENT providing a particular degree of protection against electric shock particularly regarding allowable LEAKAGE CURRENT Normal mode Single failure mode Patient leakage c...

Page 34: ...ll EMC performance of the product The product must be installed as stipulated in 1 3 2 Notice upon Installation of Product In case of issues related to EMC please follow procedures stated in 1 3 4 Countermeasures against EMC related Issues 1 3 2 Notice upon Installation of Product 1 Use either power supply cords provided or designated by GEMS or SGMS Products equipped with power source plug should...

Page 35: ...over for some reason be sure to close it before starting resuming operation Operating the system with any cover open may affect EMC performance 1 3 4 Countermeasures against EMC related Issues Generally it is very difficult to handle issues related to EMC It may be time consuming and costly General countermeasures Electromagnetic interference with other equipment 1 Electromagnetic interference may...

Page 36: ...855 W Electric Ave Milwaukee WI 53219 USA TEL 1 800 437 1171 FAX 1 414 647 4090 CANADA TEL 1 800 668 0732 LATIN SOUTH AMERICA TEL 1 305 735 2304 GE Ultrasound Europe GE Ultraschall GmbH Co KG Beethovenstrabe 239 Postfach 11 05 60 D 42655 Solingen GERMANY TEL 0130 81 6370 OLC Europe Toll Free Number 49 212 2802 207 English German Hotline 49 212 2802 208 English German French Hotline FAX 49 212 2802...

Page 37: ...Specifications 2 4 2 1 4 Recommended Ultrasound Room Layout 2 5 2 2 INSTALLATION 2 8 2 2 1 Introduction 2 8 2 2 2 Average Installation Time 2 8 2 2 3 Installation Warnings 2 8 2 2 4 Checking the Components 2 8 2 2 5 Unpacking LOGIQ α200 2 9 2 2 6 MTZ Probe Holder Installation Option 2 11 2 2 7 Transducer Connection 2 14 2 2 8 Moving into Position 2 14 2 2 9 Adjusting System Clock 2 14 2 2 10 Produ...

Page 38: ...GE MEDICAL SYSTEMS LOGIQ α200 SERVICE MANUAL 2138853 INSTALLATION 2 2 REV 0 This page is left blank intentionally ...

Page 39: ...hat you use an analyzer Dranetz Model 606 3 or Dranetz Model 626 PARAMETER LIMITS Voltage Range Korea 220 VAC 10 198 242 VAC USA 115 VAC 10 103 127 VAC Europe LA 220 240 VAC 10 198 264 VAC Japan 100 VAC 10 90 110 VAC Asia 115 VAC 10 103 127 VAC 220 240 VAC 10 198 264 VAC Voltage Range All applications MAX 500 VA Line Frequency All applications 50 60Hz 2Hz Power Transients Less than 25 of nominal p...

Page 40: ...pment may produce EMI unintentionally or as a result of a malfunction These sources include medical lasers cauterizing guns computers monitors fans gel warmers microwave ovens and cellular phones The presence of broadcast station or van may also cause interference Carefully read the following precautions before installing the unit 1 Connect the power plug for any other equipment into the fixed out...

Page 41: ...V CIRCUIT BREAKER RADIATION SHIELDING NONE REQUIRED for ULTRASOUND ENERGY TEMPERATURE 20 26 C 68 79 F for PATIENT COMFORT HUMIDITY 50 to 70 for PATIENT COMFORT HEAT DISSIPATION 2000 BTU Hr for LOGIQ α200 FLOOR LOADING Approximately 240 300 kg m2 without Accessories FLOOR CONDITION Gradient WITHIN 5 degrees LOGIQ α200 Weight 76 kg 168lbs without Accessories POWER SOURCE 230VAC 50Hz SINGLE PHASE For...

Page 42: ...xygen supply 3 Dimmer control for overhead lights 4 Film viewer 5 Film and linen storage 6 Medical equipment storage 7 Hospital grade equipment electrical outlet 8 Document storage area for operating and service manuals 9 Nearby waiting room dressing room lavatory 10 Trash bin Recommended and Alternate Ultrasound Console Floor Plans ILLUSTRATION 2 1 provides a recommended standard floor plan and a...

Page 43: ... Cabinet Dedicated power outlet Examination Table Sink GE cabinet for software and manuals 42 in Scale each square equals one square foot 61 cm 193 cm Probes PM LOGIQ 200 A 14 by 17 Recommended Floor Plan Sink Film Viewer Stool Film Supplies Linens Examination Table 193 cm 61 cm 76 cm 30 in Probes supplies GE cabinet for software and manuals Dedicated power outlet LOGIQ 200 An 8 by 10 Minimal Floo...

Page 44: ...functions occur unplug the power cord Only qualified service personnel should carry out servicing and troubleshooting Note For information regarding packing labels refer to ILLUSTRATION 2 3 LABELS ON PACKAGE 3 After being transported the unit may be very cold or hot If this is the case allow the unit to acclimate before you turn it on It requires one hour for each 2 5 C increment if it s temperatu...

Page 45: ...Lift the TOP PAD up and off 4 Remove the OPTION BOX 5 Remove the VINYL COVER 6 Remove the MIDDLE PLATE ASSY 7 Remove the MONITOR COVER 8 Lift the PACKING CASE up and off 9 Lift the Monitor up by pressing the UP DOWN button located on the Monitor Arm 10 Remove the MONITOR PAD 11 Return the Monitor arm to its lowest position 12 Carefully roll the LOGIQ α200 from the PALLET 13 Remove the Caution Labe...

Page 46: ...YSTEMS LOGIQ α200 SERVICE MANUAL 2138853 INSTALLATION 2 10 REV 0 2 2 5 Unpacking LOGIQ α200 Continued BAND PALLET CAP TOP PAD PACKING PLATE MIDDLE PLATE MONITOR PAD OPTION BOX UNPACKING LOGIQ α200 ILLUSTRATION 2 2 ...

Page 47: ...lder and two brackets are supplied with the LOGIQ α200 console as shown in ILLUSTRATION 2 4 Note Two sets of screw holes are provided at the bottom of standard probe holder and left side of keyboard to install the MTZ probe holder You can choose the most convenient position for your customer between the two sets of screw holes MTZ Probe Holder Bracket1 Right Bracket 2 Left MTZ PROBE HOLDER SET ILL...

Page 48: ...TALLATION 2 12 REV 0 2 2 6 MTZ Probe Holder Installation Option Continued 1 Assemble the MTZ probe holder at the bottom of standard probe holder by screwing four 1 4 screws as shown in ILLUSTRATION 2 5 1 2 3 4 MTZ PROBE HOLDER INSTALLATION 1 ILLUSTRATION 2 5 ...

Page 49: ...be Holder Installation Option Continued 2 After removing the Keyboard Assy Refer to 6 2 10 6 2 DISASSEMBLY RE ASSEMBLY Assemble the MTZ probe holder at the left side of keyboard by screwing four 1 4 screws as shown in ILLUSTRATION 2 6 1 2 3 4 MTZ PROBE HOLDER INSTALLATION 2 ILLUSTRATION 2 6 ...

Page 50: ...he proper rated voltage checked during preinstallation Never use a three to two prong adapter this defeats the safety ground 2 2 8 Moving into Position CAUTION Do not lift the unit by the Keyboard Do not tilt the unit more than 5 degrees to avoid tipping it over To avoid injury by tipping over Set the monitor to the lowest position before moving CAUTION Equipment Damage Possibility Lifting the con...

Page 51: ...omer Information the Product Locator Installation Card Refer to ILLUSTRATION 2 7 Mail this Installation Card Product Locator to the address corresponding to your pole Note The Product Locator Installation Card shown in ILLUSTRATION 2 7 may not be same as the provided Product Locator card PRODUCT LOCATOR INSTALLATION CARD ILLUSTRATION 2 7 ...

Page 52: ...GE MEDICAL SYSTEMS LOGIQ α200 SERVICE MANUAL 2138853 INSTALLATION 2 16 REV 0 This page is left blank intentionally ...

Page 53: ...S 3 4 3 3 1 Power Supply 3 4 3 3 2 Facility Power Receptacle 3 4 3 4 STORAGE AND OPERATION REQUIREMENTS 3 4 3 5 OPTIONAL PERIPHERALS 3 5 3 5 1 Peripherals Accessories Connector Panel 3 5 3 5 2 List of Optional Peripherals 3 8 3 5 3 Power Consumption of Optional Peripherals 3 9 3 6 TEST POINT LED DIP SWITCH AND RESET SWITCH 3 10 3 6 1 Test Point List 3 10 3 6 2 LED List 3 13 3 6 3 DIP Switch 3 13 3...

Page 54: ...GE MEDICAL SYSTEMS 2138853 LOGIQ α200 SERVICE MANUAL SYSTEM CONFIGURATION 3 2 REV 0 This page is left blank intentionally ...

Page 55: ...m configuration and specifications 3 2 DIMENSIONS Regarding LOGIQ α200 dimensions Refer to ILLUSTRATION 3 1 for planning the position of your LOGIQ α200 WEIGHT 76 kg 168 lbs NOTE LENGTH mm inches ABERRATION 5 400 15 7 350 13 8 663 26 1 335 13 2 250 9 8 245 9 6 1200 47 2 800 31 5 162 6 4 192 7 6 OVERALL DIMENSIONS ILLUSTRATION 3 1 ...

Page 56: ...ts is recommended The specific power receptacle used depends on your country s power line standards The receptacle should have International Electrotechnical Commission IEC approval or equivalent 3 4 STORAGE AND OPERATION REQUIREMENTS Probes and peripherals are shipped in separate containers Shipping weight is approximately 210 lbs 96 kg The size of the container is L82 cm x W53 cm x H140 cm 32 in...

Page 57: ...put and output connectors camera expose connector foot switch connector power connector and service tools This section indicates the pin assignment for each connector 1 4 in ILLUSTRATION 3 2 at pages 3 6 through 3 7 Note The optional ECG connector is available only for the console with the system software version 2 00 or later 1 2 3 220 240V 0 8A Max 100V 2 0A Max 115V 1 7A Max including front pan...

Page 58: ...rounded Signal ground lines are Not Isolated 1 Pin Assignment of RS232C for Service Connector Female D SUB 25 pin Pin No Signal Pin No Signal 1 Frame GND 14 2 TXD 15 3 RXD 16 4 RTS 17 5 CTS 18 6 DSR 19 7 Signal GND 20 DTR 8 DCD 21 9 22 Ring Indicator 10 23 11 24 12 5 V 1 300mA Max 25 13 1 This voltage shall be apply to this Pin in case of MODEM usage Note Output level of RS232C signals High 3V to ...

Page 59: ...nnector Pin No Output Signal Pin No Output Signal 1 FOOT SW 4 2 FOOT SW_G 5 Frame GND 3 3 Pin Assignment of Mini Jack for Controlling B W Printer Connector Stereo Mini Jack Pin No Output Signal Pin No Output Signal 1 PRINT 1 2 Signal GND 1 Printer starts printing by receiving the LOW pulse more than 100ms Note Output level of control signals indicated in the above tables are TTL level 1 2 3 4 5 ...

Page 60: ...NAME MATERIAL OF HEADSHELL AREA OF USING TYPE CATALOG NO REQUIRED ADAPTER PART NO FOR JAPAN PART NO CBF PES Abdom Convex H46022CB Not Required P9603AA P9603AD CAE PES Abdom Convex H46022CA Not Required P9603AB P9603AE MTZ PES Intercavity Convex H46022MT Not Required P9603AL P9603AU CZB NORYL Neonatal Convex H45202CZ Not Required 2152402 2152422 CS PES Abdom Convex H45222CS Not Required 2202315 220...

Page 61: ...ABLES TABLE 3 4 LIST OF CONNECTING CABLES NAME PART NO FIGURE NOTE Printer Install Kit 2176459 Printer Cable Assy BNC cable Shutter cable For B W Printer 3 5 3 Power Consumption of Optional Peripherals The table below shows the power consumption of each optional peripheral for LOGIQ α200 TABLE 3 5 POWER CONSUMPTION OF OPTIONAL RECORDING DEVICES DEVICE MODEL POWER CONSUMPTION Video Cassette Recorde...

Page 62: ...ine No 8 TP9 5V for analog digital TP10 Input echo signal from the ESP Assy TP V5_1 5V for digital Edge of board MST ASSY TP601 Gnd Edge of board TP602 5V for digital Edge of board FOOT S W Foot switch status CLK2_25M 25MHz clock CLK 9M 9MHz clock CLK36M 36MHz clock RTC ASSY 5V 5V for digital Edge of board 15V 15V for analog Edge of board 15V 15V for analog Edge of board THV High voltage for trans...

Page 63: ... of board V 5 5V for analog digital Edge of board V5 5V for digital Edge of board CDA H High position of aperture reference voltage Edge of board CDA L Low position of aperture reference voltage Edge of board CDA ADJ For Adjusting aperture refer ence voltage CDFT OUT Positive control voltage of CDF CDFN OUT Negative control voltage of CDF CDF ADJ For adjusting CDF control cir cuit TFC OUT Output o...

Page 64: ...N 3 12 REV 0 3 6 1 Test Point List continued TABLE 3 8 TEST POINT LOCATION NAME DESCRIPTION POSITION ESP ASSY TP3 GND TP6 GND V15 15V for Analog CONN ASSY 5V 5V for digital Edge of board 15V 15V for analog Edge of board SHV High voltage for high voltage switch Edge of board ...

Page 65: ...of board DS5 DSP running Edge of board ON OFF HV ASSY DS1 HV ASSY status ON normal Edge of board ON OFF SMPS ASSY GREEN LED SMPS ASSY status ON nor mal SMPS ASSY ON OFF Note Three LED DS2 DS4 should be blinked when the system operated with the Probe 3 6 3 DIP Switch The table below shows the DIP Switch list for LOGIQ α200 TABLE 3 8 DIP SWITCH Setting LOCATION NAME Switch No DESCRIPTION POSITION MS...

Page 66: ...GE MEDICAL SYSTEMS 2138853 LOGIQ α200 SERVICE MANUAL SYSTEM CONFIGURATION 3 14 REV 0 This page is left blank intentionally ...

Page 67: ...APTER 4 FUNCTIONAL CHECKS TABLE OF CONTENTS SECTION TITLE PAGE 4 1 INTRODUCTION 4 3 4 1 1 Required Equipment 4 3 4 2 FUNCTIONAL CHECK PROCEDURES 4 3 4 2 1 Basic Controls 4 3 4 2 2 M Mode Check 4 5 4 3 SMPS ADJUSTMENTS 4 7 4 3 1 SMPS Assy Access 4 7 4 3 2 SMPS Adjustment Procedure 4 9 ...

Page 68: ...GE MEDICAL SYSTEMS LOGIQ 200 SERVICE MANUAL 2138853 FUNCTIONAL CHECKS 4 2 REV 0 This page is left blank intentionally ...

Page 69: ...elect another gray scale Map The gray scale adjusts to each new Map selected 5 Press Dyn Range Arrow up or down key At lower Dynamic Range settings image speckle fades and prominent objects in the display are more pronounced from the background image 6 Rotate Depth knob The depth of image should be increased decreased 7 Slide TGC potentiometers pots Image grows darker or brighter at depth equivale...

Page 70: ...ANUAL 2138853 FUNCTIONAL CHECKS 4 4 REV 3 4 2 1 Basic Controls Continued ÇÇ ÇÇ ÇÇ ÇÇ ÇÇ ÇÇ ÇÇ ÇÇ ÇÇ ÇÇ ÇÇ ÇÇ ÇÇ ÇÇ ÇÇ 0 cm 1 cm 5 cm 10 cm ÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇ Map1 Cine Gauge FOV GE FA EE Category B MODE DISPLAY SCREEN ILLUSTRATION 4 1 ...

Page 71: ...e CRT monitor 14 Rotate B M Gain knob The M timeline should get brighter with CW rotation and darker with CCW 15 Press Dyn Range Arrow up or down key Dynamic Range affects grays and the last added scan mode to adjust the basic B M must be off Turn Dynamic Range down to increase contrast turn up to soften 16 Press Sweep Speed key Press it again to exit The timeline speed should increase to 4 second...

Page 72: ... ÇÇ ÇÇ ÇÇ ÇÇ ÇÇ ÇÇ ÇÇ Map1 FOV ÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇÇ ÇÇ ÇÇ ÇÇ ÇÇ ÇÇ ÇÇ ÇÇ ÇÇ ÇÇ ÇÇ ÇÇ ÇÇ ÇÇ M GE Cine Gauge Cine Gauge FA EE Category M MODE DISPLAY SCREEN ILLUSTRATION 4 2 Note You can select several types of display formats by using the Setup Menu For the Preset Menu refer to Customizing Your System in the LOGIQ 200 User Manual ...

Page 73: ...outlet conforms to the proper power line standards Refer to Chapter 2 Installation Note If the adjustment pot is turned to far clockwise the SMPS output shuts down to protect the circuits against the over voltage In that case power the LOGIQ 200 OFF and turn the pot counterclockwise all the way Then power it ON and try to adjust the SMPS again The SMPS Assy is in the bottom of the LOGIQ 200 as sho...

Page 74: ... the four screw caps and unscrew the screws to remove the Right Cover as shown in the ILLUSTRATION 4 4 Right Cover RIGHT COVER REMOVAL ILLUSTRATION 4 4 3 Remove the four screw caps and unscrew the screws to remove the Rear Cover as shown in the ILLUSTRATION 4 5 ÉÉ ÉÉ ÉÉ ÉÉ ÉÉ Rear Cover REAR COVER REMOVAL ILLUSTRATION 4 5 ...

Page 75: ...15V on ESP BD ASSY TP3 Ground on ESP BD ASSY 1 on SMPS ASSY See ILLUSTRATION 4 6 5V for Analog Digital V 5 5V on ESP BD ASSY TP3 Ground on ESP BD ASSY 2 on SMPS ASSY See ILLUSTRATION 4 6 12 for Monitor B 12V on Monitor Input Connector GND on Monitor Input Connector 3 on SMPS ASSY See ILLUSTRATION 4 6 5V for Digital TP 602 5V on MST BD ASSY TP 601 GND on MST BD ASSY 4 on SMPS ASSY See ILLUSTRATION ...

Page 76: ...FUNCTIONAL CHECKS 4 10 REV 0 4 3 2 SMPS Adjustment Procedure Continued ÉÉÉ ÉÉÉ ÉÉÉ ÉÉÉ ÉÉÉ 1 Adjustment Point for 5V 15V SMPS 2 Adjustment Point for 5V SMPS 3 Adjustment Point for 12V SMPS 4 Adjustment Point for 5V SMPS 1 2 3 4 SMPS ASSY ILLUSTRATION 4 6 ...

Page 77: ...200 SERVICE MANUAL 2138853 5 1 DIAGRAMS REV 0 CHAPTER 5 DIAGRAMS TABLE OF CONTENTS SECTION TITLE PAGE 5 1 INTRODUCTION 5 3 5 2 LOGIQ 200 SYSTEM 5 3 5 3 BLOCK DIAGRAM 5 4 5 4 WIRING DIAGRAM 5 5 5 5 CIRCUIT BOARD DESCRIPTION 5 7 ...

Page 78: ...GE MEDICAL SYSTEMS LOGIQ 200 SERVICE MANUAL 2138853 5 2 DIAGRAMS REV 0 This page is left blank intentionally ...

Page 79: ...echnology were used to increase console ease of use System Features The key design goals of this system are High Image Quality Increased User Productivity Multiple Clinical Applications Planned Upgradeability High Mobility Types of Applications The system supports many clinical uses Scan and display parameters may be user selected to default to desired values for each application The system preset...

Page 80: ...IAGRAMS REV 3 5 3 BLOCK DIAGRAM DSC B W Monitor MST Keyboard Rear Panel Foot Switch Power Supply RTC CONN HV ÉÉ ÉÉ ÉÉÉ ÉÉÉ ESP Shutter Control NOVA BUS RTC BUS To Mother board To Mother board To Mother board ECG LOGIQ 200 SYSTEM BLOCK DIAGRAM ILLUSTRATION 5 1 ...

Page 81: ...100 115 0 110 POWER SW L Brown N Light Blue G Green Yellow 4 1 3 2 4 NFB1 Core 10A 115V 5A 220V L N TB1 TB2 TB2 NFB3 NFB4 2A 115V 1A 220V NFB2 3A 9 8 7 12 11 10 4 5 6 2 3 16 17 15 ISOLATION TRANSFORMER TB3 L N G N L G 1 2 1 2 N C 5 VD 5V D G N D G N D 15 VA 15 VA 5 VA G N D G N D 5V G N D 12 V G N D 12 V G N D J10 J13 J11 J28 J29 G N D 1 2 E X T V I D I N R T N F O O T S W S H T R G F O O T S W G ...

Page 82: ...15V 100V 0V 115V 100V 0V 115V 100V 0V 115V 100V 0V Light Blue 12 11 10 9 8 7 6 5 4 3 2 1 115V 100V 0V 115V 100V 0V 115V 100V 0V 115V 100V 0V Light Blue 100 V 115 V 220 240 V INPUT 220 240 V OUTPUT 115 V TB2 TB2 TB2 TB2 AC IN Peripheral Out AC IN Peripheral Out AC IN Peripheral Out AC IN Peripheral Out Brown Brown Brown Orange Orange Orange Orange Gray Gray Gray Gray Brown Light Blue Orange Gray Br...

Page 83: ...oard of the LOGIQ 200 system TABLE 5 1 CIRCUIT BOARD DESCRIPTION CARD CAGE SLOT BOARD NAME DESCRIPTION NOTE 1 ESP BD ASSY Echo Signal Processor Board Assembly 2 RTC BD ASSY Real Time Controller Board Assembly 3 DSC BD ASSY Digital Scan Convertor Board Assembly 4 MST BD ASSY Master Board Assembly CONN BD ASSY Connector Board Assembly HV BD ASSY High Voltage Board Assembly MOTHER ASSY Motherboard As...

Page 84: ...GE MEDICAL SYSTEMS LOGIQ 200 SERVICE MANUAL 2138853 5 8 DIAGRAMS REV 0 This page is left blank intentionally ...

Page 85: ...No 205 6 38 6 2 12 TGC Assy FRU No 206 6 39 6 2 13 Encoder Set FRU No 207 6 40 6 2 14 Freeze Record Key Assy FRU No 208 6 41 6 2 15 TGC Knob Set FRU No 209 6 42 6 2 16 Probe Protector Sheet Set FRU No 210 6 43 6 2 17 Key Sheet Assy FRU No 211 6 44 6 2 17 Key Sheet Assy continued FRU No 211 6 45 6 2 18 P C Board s FRU No 301 through 304 6 46 6 2 19 HV Assy FRU No 305 6 48 6 2 20 CONN Assy FRU No 30...

Page 86: ...16 6 84 6 2 39 AC Fan Assy FRU No 417 6 86 6 2 40 Front Caster FRU No 418 6 88 6 2 41 Rear Caster FRU No 419 6 89 6 2 42 Bumper Set FRU No 420 6 90 6 2 43 EMI Cover L FRU No 421 6 91 6 2 44 PCB Guide Assy FRU No 422 6 92 6 2 45 Terminal Block 12 Assy FRU No 501 6 94 6 2 46 Terminal Block 6 Assy FRU No 502 6 96 6 2 47 SMPS Assy FRU No 503 6 98 6 2 48 Power Trans Assy FRU No 504 6 100 6 2 49 Circuit...

Page 87: ...157730 1 115V NTSC 2 SWIVEL VER0 3 OPERATOR CONSOLE ASSY 2157731 1 220V PAL 2 SWIVEL VER0 3 OPERATOR CONSOLE ASSY 2157732 1 220V NTSC 2 SWIVEL VER0 3 CABLE HOOK ARM 2170597 1 1 1 1 1 1 1 1 1 MTZ HOLDER ASSY 2175435 RIGHT MTZ HOLDER ASSY 2175436 LEFT FOOT SWITCH ASSY 2162242 CINE 2174971 1 1 1 1 1 PRINTER INSTALL KIT 2176459 1 1 1 1 1 1 1 1 1 OPERATOR MANUAL 2138852 100 1 1 1 1 1 1 1 English SERVIC...

Page 88: ... ASSY 2184520 1 220V PAL 2 SWIVEL VER1 1 OPERATOR CONSOLE ASSY 2184521 1 220V NTSC 2 SWIVEL VER1 1 CABLE HOOK ARM 2170597 1 1 1 1 1 1 1 1 1 Include hook arm space MTZ HOLDER ASSY 2175435 RIGHT MTZ HOLDER ASSY 2175436 LEFT FOOT SWITCH ASSY 2162242 1 1 1 CINE 2174971 1 1 1 1 1 1 PRINTER INSTALL KIT 2176459 1 1 1 1 1 1 1 1 1 OPERATOR MANUAL 2183922 100 1 1 1 1 1 1 1 English SERVICE MANUAL 2138853 1 1...

Page 89: ...205675 1 115V NTSC 2SW VER 2 01 CONSOLE ASSY 2205676 1 220V PAL 2SW VER 2 01 CONSOLE ASSY 2205677 1 220V NTSC 2SW VER 2 01 PROBE CABLE ARM 2215992 1 1 1 1 2 0 UPGRADE KIT 2217555 Include CINE 2 0 UPGRADE KIT 2211832 CABLE HOOK ARM 2170597 1 1 1 1 1 1 1 1 1 1 1 Include hook arm space MTZ HOLDER ASSY 2175435 RIGHT MTZ HOLDER ASSY 2175436 LEFT FOOT SWITCH ASSY 2162242 1 1 1 CINE 2174971 1 1 1 1 1 1 1...

Page 90: ...GE MEDICAL SYSTEMS LOGIQ α200 SERVICE MANUAL 2138853 6 6 RENEWAL PARTS REV 5 This page is left blank intentionally ...

Page 91: ...V 4 OPERATOR CONSOLE ASSY MONITOR BLOCK FRU 100 105 See ILLUSTRATION 6 2 KEYBOARD BLOCK FRU 201 204 See ILLUSTRATION 6 3 PC BOARDS 2 See ILLUSTRATION 6 4 BODY BLOCK See ILLUSTRATION 6 5 PC BOARDS 1 See ILLUSTRATION 6 4 POWER SUPPLY BLOCK See ILLUSTRATION 6 6 ILLUSTRATION 6 1 ...

Page 92: ...GE MEDICAL SYSTEMS LOGIQ α200 SERVICE MANUAL 2138853 6 8 RENEWAL PARTS REV 4 OPERATOR CONSOLE ASSY 1 7 102 101 103 104 105 106 ILLUSTRATION 6 2 ...

Page 93: ... 1 MONITOR ASSY NTSC 2167605 2 1 1 Including Caution Label Glass Filter For America USA Chile 6 2 1 MONITOR ASSY PAL 2167606 2 1 1 Including Caution Label Glass Filter For America Brazil 6 2 1 101 POT SET ASSY 2148193 1 2 Bright and Contrast Include Knob 6 2 2 102 CRT FILTER PLASTIC 2175437 1 2 Include 2 brackets For Japan Taiwan Phillippines China Hongkong India Southeast Asia Australia Europe 6 ...

Page 94: ...GE MEDICAL SYSTEMS LOGIQ α200 SERVICE MANUAL 2138853 6 10 RENEWAL PARTS REV 5 OPERATOR CONSOLE ASSY 2 7 204 203 202 201 PG Rivet 207 208 205 206 209 210 211 ILLUSTRATION 6 3 ...

Page 95: ...75439 1 2 Include Probe Cup Holder and Probe Sheet 6 2 7 202 HOLDER BRACKET ASSY 2148199 1 2 Include PG Rivet 6 2 8 203 KEYBOARD COVER SET 2148200 1 2 Top Bottom with Holder Brkt In 6 2 9 204 KEYBOARD ASSY 2158943 1 1 1 Include FRU 205 208 6 2 10 2158943 2 1 1 For S W Version 2 0 6 2 10 205 TRACKBALL ASSY 2174984 1 1 6 2 11 206 TGC ASSY 2174985 1 1 6 2 12 207 ENCODER SET 2174986 1 1 6 2 13 208 FRE...

Page 96: ...GE MEDICAL SYSTEMS LOGIQ α200 SERVICE MANUAL 2138853 6 12 RENEWAL PARTS REV 4 OPERATOR CONSOLE ASSY 3 7 301 304 305 306 307 308 309 ILLUSTRATION 6 4 ...

Page 97: ... software 6 2 18 302 DSC ASSY 2148203 2 1 1 Version 1 2 software 6 2 18 2199767 3 1 1 Version 2 01 software 6 2 18 2243263 1 1 Version 1 2 software including Cine 6 2 18 2243262 2 1 1 Version 2 01 software including Cine 6 2 18 303 RTC ASSY 2185498 1 1 Version 1 2 software 6 2 18 2218506 1 1 Version 2 0 software 6 2 18 2218506 2 1 1 Version 2 01 software 6 2 18 304 ESP ASSY 2148205 3 1 1 Version 1...

Page 98: ...MEDICAL SYSTEMS LOGIQ α200 SERVICE MANUAL 2138853 6 14 RENEWAL PARTS REV 4 OPERATOR CONSOLE ASSY 4 7 401 402 403 404 405 406 407 408 409 410 411 412 413 414 415 416 417 418 419 420 421 422 ILLUSTRATION 6 5 ...

Page 99: ...IGHT COVER 2158053 1 1 6 2 28 406 FRONT BASE COVER 2148217 1 1 6 2 29 407 FRONT COVER 2148218 1 1 6 2 30 408 REAR COVER 2175441 1 2 Include Caution Label 6 2 31 409 TOP COVER 2158086 1 1 6 2 32 410 PRINTER COVER 2148220 1 2 6 2 32 411 PRINTER BRACKET ASSY 2167717 1 2 For Korea Taiwan Phillippines USA Chile 6 2 34 PRINTER BRACKET ASSY 2167718 1 2 For Japan 6 2 34 PRINTER BRACKET ASSY 2167719 1 2 Fo...

Page 100: ...wan Phillippines USA Chile 6 2 37 REAR PANEL ASSY 2167620 1 1 For Japan 6 2 37 REAR PANEL ASSY 2167621 1 1 For China Hongkong Southeast Asia India Auatralia Europe Brazil 6 2 37 416 POWER S W ASSY 2148226 1 1 6 2 38 417 AC FAN ASSY 2175652 1 1 Ver 0 3 Ver 1 1 6 2 39 2206020 Ver 2 0 2245589 Ver 2 01 418 FRONT CASTER 2148228 2 1 6 2 40 2192111 419 REAR CASTER 2148051 2 1 Swivel For Japan Taiwan Phil...

Page 101: ...GE MEDICAL SYSTEMS LOGIQ α200 SERVICE MANUAL 2138853 6 17 RENEWAL PARTS REV 4 This page is left blank intentionally ...

Page 102: ...GE MEDICAL SYSTEMS LOGIQ α200 SERVICE MANUAL 2138853 6 18 RENEWAL PARTS REV 4 OPERATOR CONSOLE ASSY 5 7 ÉÉÉ ÉÉÉ ÉÉÉ ÉÉÉ ÉÉÉ ÉÉÉ 503 504 505 502 501 ILLUSTRATION 6 6 ...

Page 103: ...6 2 46 503 SMPS ASSY 2148233 1 1 6 2 47 504 POWER TRANS ASSY INPUT 220V OUTPUT 115V 2167617 1 1 For Korea Chile 6 2 48 POWER TRANS ASSY INPUT 100V OUTPUT 100V 2167618 1 1 For Japan 6 2 48 POWER TRANS ASSY INPUT 115V OUTPUT 115V 2170698 1 1 For Taiwan Phillippines USA 6 2 48 POWER TRANS ASSY INPUT 220V OUTPUT 220V 2170699 1 1 For China Hongkong India Southeast Asia Australia Europe Brazil 6 2 48 50...

Page 104: ... 1 1A 250V 500mA 250V 2A 125V 6 3 602 FLAT CABLE SET 2148237 1 1 Cable Assy 49 51 603 CABLES SET 2148238 1 1 Cable Assy 2 3 5 6 7 8 9 10 11 12 13 14 15 16 17 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 37 39 40 41 42 43 44 45 46 47 48 61 62 63 64 65 604 HARDWARE SET 2148239 1 2 2157994 2166804 4 2157951 12 2158063 2157971 2157973 2158077 2 2160325 2159642 10 2159634 10 2159625 10 2162241 10 21731...

Page 105: ... PROBE P9603AU 1 1 Not Include MTZ HOLDER ASSY P9603AL 1 1 Not Include MTZ HOLDER ASSY For Japan Only 704 CZB PROBE 2152422 1 1 2152402 1 1 For Japan Only 705 LH PROBE P9601AS 1 1 P9601AC 1 1 For Japan Only 706 LE PROBE P9601AR 1 1 P9601AB 1 1 For Japan Only 707 LI PROBE P9601AW 1 1 P9601AG 1 1 For Japan Only 708 LT PROBE P9601AX 1 1 P9601AJ 1 1 For Japan Only 709 LB PROBE P9601AQ 1 1 P9601AA 1 1 ...

Page 106: ...GE MEDICAL SYSTEMS LOGIQ α200 SERVICE MANUAL 2138853 6 22 RENEWAL PARTS REV 4 This page is left blank intentionally ...

Page 107: ...S INSIDE BECOME THOROUGHLY FAMILIAR WITH ALL HAZARDOUS VOLTAGES AND HIGH CURRENT LEVELS TO AVOID ACCIDENTAL CONTACT CAUTION Do not wear the ESD wrist strap when you remove the SMPS Assy Turn OFF power and unplug the power cord before removing any part of SMPS Assy However be sure to turn off power and wear the strap before you remove a circuit boards CAUTION Do NOT unplug the power cord before tur...

Page 108: ...ELEASING ARM WHEN MONITOR IS NOT INSTALLED WILL CAUSE SUDDEN UPWARD MOVEMENT KEEP YOUR HEAD AND BODY AWAY FROM ABOVE VIDEO MONITOR SUPPORT ARM Refer to ILLUSTRATION 6 7 1 Turn OFF the system 2 Lift and set the Monitor to the highest position by pushing down the Up down Release Button located on the Swing Arm Assy 3 Remove the Monitor Bottom Cover Refer to 6 2 4 4 Disconnect the connector 5 Unscrew...

Page 109: ...DICAL SYSTEMS LOGIQ α200 SERVICE MANUAL 2138853 6 25 RENEWAL PARTS REV 4 6 2 1 Monitor Assy FRU No 100 Continued 100 Monitor Assy connector 1 ÉÉÉ ÉÉÉ ÉÉÉ ÉÉÉ ÉÉÉ ÉÉÉ MONITOR ASSY DISASSEMBLY ILLUSTRATION 6 7 ...

Page 110: ...r Soldering Iron Procedure Refer to ILLUSTRATION 6 8 1 Turn OFF the system 2 Remove the Monitor Cover Set Refer to 6 2 4 3 Remove two Potientiometer Knob and unscrew two hexagonal nuts on each Pot 4 Remove the wire from PCB with Soldering Iron 5 Remove the Pot Set Assy 101 Pot Set Assy Hexagonal Nut Potientiometer Knob POT SET DISASSEMBLY ILLUSTRATION 6 8 ...

Page 111: ...Time Required 2 Minutes Tool Required Not necessary Procedure Refer to ILLUSTRATION 6 9 1 Turn OFF the system 2 Push the two Filter Clamp Assy inward 1 as shown in ILLUSTRATION 6 9 and remove Filter Clamp Assy 3 Pull forward CRT Filter 102 CRT Filter 106 Clamp Set 1 1 CRT FILTER DISASSEMBLY ILLUSTRATION 6 9 ...

Page 112: ... the Monitor to the highest position by pushing down the Up down Release Button located on the Swing Arm Assy 3 Unscrew one screws 1 and remove the Monitor Bottom Cover 4 Unscrew four screws 2 5 and remove the Monitor Rear Cover 5 Unscrew four screws 6 9 and remove the Monitor Front Cover 1 2 3 4 5 6 7 8 103 Monitor Cover Set Monitor Front Cover Monitor Bottom Cover Monitor Rear Cover 9 MONITOR CO...

Page 113: ...nutes Tool Required Screwdriver Procedure Refer to ILLUSTRATION 6 11 1 Turn OFF the system 2 Remove the Monitor Assy FRU No 100 from the Swing Arm Assy Refer to 6 2 1 3 Remove the Monitor Rear Cover Refer to 6 2 4 4 Unscrew four screws 1 4 5 Remove the Tilt Assy 104 Tilt Assy 1 2 3 4 TILT ASSY DISASSEMBLY ILLUSTRATION 6 11 ...

Page 114: ...e FRU No 105 Time Required 6 Minutes Tool Required Screwdriver Procedure Refer to ILLUSTRATION 6 12 1 Turn OFF the system 2 Remove the Monitor Assy FRU No 100 from the Swing Arm Assy Refer to 6 2 1 3 Remove the Monitor Space Plate 105 Monitor Space Plate MONITOR SPACE PLATE DISASSEMBLY ILLUSTRATION 6 12 ...

Page 115: ...No 201 Time Required 2 Minutes Tool Required Not necessary Procedure Refer to ILLUSTRATION 6 13 1 Turn OFF the system 2 Pull the PG latch located at bottom of the Probe Holder 3 Pull the Probe Holder out as shown in ILLUSTRATION 6 13 PG latch 201 Probe Holder PROBE HOLDER DISASSEMBLY ILLUSTRATION 6 13 ...

Page 116: ... No 202 Time Required 3 Minutes Tool Required Screwdriver Procedure Refer to ILLUSTRATION 6 14 1 Turn OFF the system 2 Remove the Probe Holder FRU No 201 Refer to 6 2 7 3 Unscrew three screws 1 3 4 Remove the Holder Bracket Assy 1 2 3 202 Holder Bracket Assy HOLDER BRACKET ASSY DISASSEMBLY ILLUSTRATION 6 14 ...

Page 117: ...GE MEDICAL SYSTEMS LOGIQ α200 SERVICE MANUAL 2138853 6 33 RENEWAL PARTS REV 4 This page is left blank intentionally ...

Page 118: ...Holder Bracket Assy FRU No 202 Refer to 6 2 8 4 Unscrew two ground screws 1 and 2 on the front bottom of Keyboard 5 Unscrew three screws on the PCB that hold Ground Terminal wire and remove three clamps 6 Unscrew two screws and remove two Ground Terminals 7 Disconnect the connector and cut the tie wraps off 8 Remove the KB Cover Top Assy with Keyboard Assy FRU No 204 9 Unscrew eight screws 3 10 an...

Page 119: ...L PARTS REV 4 6 2 9 Keyboard Cover Set FRU No 203 Continued Connector Ground Terminal 11 12 14 15 17 18 19 13 16 KB Cover Bottom Keyboard Assy with KB Cover Top Assy Bottom View Keyboard Cover Set 3 4 5 6 7 8 9 10 1 2 KEYBOARD COVER SET DISASSEMBLY ILLUSTRATION 6 15 ...

Page 120: ...16 1 Turn OFF the system 2 Unscrew two screws 1 and 2 on the front bottom of Keyboard 3 Unscrew three screws on the PCB that hold Ground Terminal wire and remove three clamps 4 Unscrew two ground screws and remove two Ground Terminals 5 Disconnect the connector and cut the tie wraps off 6 Pull the Keyboard Assy out with KB Cover Top Assy 7 Unscrew eight screws 3 10 and remove KB Cover Top Assy ...

Page 121: ...MANUAL 2138853 6 37 RENEWAL PARTS REV 4 6 2 10 Keyboard Assy FRU No 204 Continued 1 2 204 Keyboard Assy Connector Ground Terminals Keyboard Assy with KB Cover Top Assy Bottom View 3 4 5 6 7 8 9 10 KEYBOARD ASSY DISASSEMBLY ILLUSTRATION 6 16 ...

Page 122: ...8 Minutes Tool Required Screwdriver Procedure Refer to ILLUSTRATION 6 17 1 Turn OFF the system 2 Pull the Keyboard Assy out with KB Cover Top Assy Refer to 6 2 10 3 Unscrew five screws 1 5 and disconnect one connect on Trackball Assy 4 Remove Trackball Assy Connector 5 1 2 4 3 CN1 TRACKBALL ASSY DISASSEMBLY ILLUSTRATION 6 17 ...

Page 123: ...wdriver Procedure Refer to ILLUSTRATION 6 18 1 Turn OFF the system 2 Pull the Keyboard Assy out with KB Cover Top Assy Refer to 6 2 10 3 Pull out the TGC Knobs 5 12 on Keyboard Assy 4 Unscrew four screws 1 4 and disconnect one connect CN2 5 Remove TGC Assy 1 2 4 3 CN1 CN2 CN6 5 12 TGC Knob 6 7 8 9 10 11 KEYBOARD ASSY DISASSEMBLY ILLUSTRATION 6 18 ...

Page 124: ... Procedure Refer to ILLUSTRATION 6 19 1 Turn OFF the system 2 Pull the Keyboard Assy out with KB Cover Top Assy Refer to 6 2 10 3 Unscrew hexagonal screws 1 on each Encoder Knob and remove three knobs 4 Unscrew four screws on each Encoder Bracket and disconnect three connects CN2 CN4 5 Remove Encoder Set CN2 CN4 CN3 Encoder Bracket 1 ENCODER SET DISASSEMBLY ILLUSTRATION 6 19 ...

Page 125: ...quired 10 Minutes Tool Required Screwdriver Procedure Refer to ILLUSTRATION 6 20 1 Turn OFF the system 2 Pull the Keyboard Assy out with KB Cover Top Assy Refer to 6 2 10 3 Unscrew five screws 1 5 and disconnect the connect 4 Remove Freeze Record Key Assy 1 2 Connector 3 4 5 FREEZE RECORD KEY ASSY DISASSEMBLY ILLUSTRATION 6 20 ...

Page 126: ...8853 6 42 RENEWAL PARTS REV 4 6 2 15 TGC Knob Set FRU No 209 Time Required 5 Minutes Tool Required None Procedure Refer to ILLUSTRATION 6 21 1 Turn OFF the system 2 Pull out the TGC Knobs 1 8 1 2 3 4 5 6 7 8 TGC KNOB SET DISASSEMBLY ILLUSTRATION 6 21 ...

Page 127: ...43 RENEWAL PARTS REV 5 6 2 16 Probe Protector Sheet Set FRU No 210 Time Required 1 Minutes Tool Required None Procedure Refer to ILLUSTRATION 6 22 1 Turn OFF the system 2 Insert the Probe Protector Sheets 1 2 1 2 PROBE PROTECTOR SHEET SET ILLUSTRATION 6 22 ...

Page 128: ...hexagonial wrench remove the three Encoder knobs Refer to 6 2 13 5 Remove the Freeze Record key Refer to 6 2 14 6 Disconnect five connectors CN1 CN5 7 Unscrew 4 screws 16 19 and disconnect the connector CN6 in the the circuit board A from the circuit board B 8 Using the boxer 5mm unscrew four studs ST1 ST4 9 Unscrew fifteen screws 1 15 and remove the circuit board B from the keyboard bracket 10 Un...

Page 129: ...et Assy continued FRU No 211 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 A B 20 21 22 23 24 25 26 27 28 Keyboard Bracket 30 TGC Track Ball CN1 CN2 CN3 CN4 CN5 ST1 ST2 ST4 ST3 B Track Ball TGC 31 211 KeySheet Assy A CN6 ST5 204 Keyboard Assy CN6 KEY SHEET ASSY DISASSEMBLY ILLUSTRATION 6 23 ...

Page 130: ... when mounting dismounting each board Surface mount IC boards are very susceptible to damage from flex torque Make sure the Dip switch position is set correctly before changing the MST Assy Refer to chapter 3 section 3 6 3 DIP Switch for correct setting Refer to ILLUSTRATION 6 24 1 Turn OFF the system 2 Remove the Right Cover FRU 405 Refer to 6 2 28 3 Remove the PCB Guide Assy FRU 422 Refer to 6 2...

Page 131: ...S LOGIQ α200 SERVICE MANUAL 2138853 6 47 RENEWAL PARTS REV 5 6 2 18 P C Board s FRU No 301 through 304 Continued a a b Bracket Nest Cardcage No Board 1 ESP 2 RTC 3 DSC 4 MST 1 2 3 4 P C BOARD S REMOVAL ILLUSTRATION 6 24 ...

Page 132: ...RU 404 Refer to 6 2 27 3 Remove the Right Cover FRU 405 Refer to 6 2 28 4 Remove the Front Base Cover FRU 406 Refer to 6 2 29 5 Remove the Front Cover FRU 407 Refer to 6 2 30 6 Unscrew four screws 1 4 and remove transparent plastic cover 7 Remove the EMI Core 8 8 Unscrew three screws 5 7 being used fixing TR 9 Disconnect a connector J31 and connector J12 on the Mother Assy 10 Unscrew and remove fo...

Page 133: ...SYSTEMS LOGIQ α200 SERVICE MANUAL 2138853 6 49 RENEWAL PARTS REV 5 6 2 19 HV Assy FRU No 305 Continued 1 2 3 4 S1 S2 S3 S4 J31 J30 5 6 7 Transparent Plastic Cover 305 HV Assy HV ASSY DISASSEMBLY ILLUSTRATION 6 25 ...

Page 134: ...13 Minutes Tool Required Screwdriver Procedure Refer to ILLUSTRATION 6 26 1 Turn OFF the system and plug out 2 Remove the Shield Panel FRU 307 Refer to 6 2 21 3 Unscrew four screws 1 5 4 Disconnect two connector 5 Pull out the Probe CONN Assy 306 CONN Assy Two Connectors 1 2 3 4 5 CONN ASSY DISASSEMBLY ILLUSTRATION 6 26 ...

Page 135: ...GE MEDICAL SYSTEMS LOGIQ α200 SERVICE MANUAL 2138853 6 51 RENEWAL PARTS REV 5 This page is left blank intentionally ...

Page 136: ...ord before OFF the power switch Refer to ILLUSTRATION 6 27 1 Turn OFF the system plug out 2 Remove the Left Cover FRU 404 Refer to 6 2 27 3 Remove the Right Cover FRU 405 Refer to 6 2 28 4 Remove the Front Base Cover FRU 406 Refer to 6 2 29 5 Remove the Front Cover FRU 407 Refer to 6 2 30 6 Remove the Power S W Assy FRU 416 Refer to 6 2 38 7 Unscrew eleven screws 1 11 8 Remove the Shield Panel FRU...

Page 137: ...GE MEDICAL SYSTEMS LOGIQ α200 SERVICE MANUAL 2138853 6 53 RENEWAL PARTS REV 5 6 2 21 Shield Panel FRU No 307 Continued 1 2 3 4 5 6 8 9 10 11 7 307 Shield Panel SHIELD PANEL DISASSEMBLY ILLUSTRATION 6 27 ...

Page 138: ... torque Refer to ILLUSTRATION 6 28 1 Turn OFF the system 2 Remove the Left Cover FRU 404 Refer to 6 2 27 3 Remove the Right Cover FRU 405 Refer to 6 2 28 4 Remove the Front Base Cover FRU 406 Refer to 6 2 29 5 Remove the Front Cover FRU 407 Refer to 6 2 30 6 Remove the EMI Cover L FRU 421 Refer to 6 2 43 7 Remove the four PCB Guide Assy FRU 422 Refer to 6 2 44 8 Remove the four P C Boards FRU 301 ...

Page 139: ...RENEWAL PARTS REV 5 6 2 22 Mother Assy FRU No 308 Continued 308 Mother Assy 1 2 3 4 5 6 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 J1 J2 J3 J4 J5 J6 J7 J8 J9 J10 J11 J12 J13 7 a b Bracket Harness J28 J29 NEST MOTHER ASSY DISASSEMBLY ILLUSTRATION 6 28 ...

Page 140: ...nting each board Surface mount IC boards are very susceptible to damage from flex torque Procedure Refer to ILLUSTRATION 6 29 1 Turn OFF the system 2 Remove the Left Cover FRU 404 Refer to 6 2 27 3 Remove the Right Cover FRU 405 Refer to 6 2 28 4 Remove the Front Base Cover FRU 406 Refer to 6 2 29 5 Remove the Front Cover FRU 407 Refer to 6 2 30 6 Remove the Rear Cover FRU 408 Refer to 6 2 31 7 Re...

Page 141: ...No 309 Continued 11 Remove the CONN Assy FRU 306 Refer to 6 2 20 12 Remove the Mother Assy FRU 308 Refer to 6 2 22 13 Remove the eight Screws 1 8 on both side of the Nest Box Refer to ILLUSTRATION 6 29 14 Remove the Nest Box FRU 309 309 Nest Box 1 2 3 4 5 6 7 8 NEST BOX DISASSEMBLY ILLUSTRATION 6 29 ...

Page 142: ...ug the power cord from outlet 2 Remove the Monitor Assy FRU No 100 from the Swing Arm Assy Refer to 6 2 1 3 Remove the Monitor Space Plate FRU No 105 Refer to 6 2 6 4 Unscrew four screws 1 4 and remove the Cover located at the bottom of the Swing Arm Assy 5 Unscrew one screw to remove Ground Terminal 6 Unscrew two supporters 5 and 6 and two screws 7 and 8 7 Remove the Guide 8 Unscrew one screws 10...

Page 143: ...GE MEDICAL SYSTEMS LOGIQ α200 SERVICE MANUAL 2138853 6 59 RENEWAL PARTS REV 5 6 2 24 Swing Arm Assy FRU No 401 Continued 1 2 3 4 9 401 Swing Arm Assy 6 5 7 8 NECK ASSY DISASSEMBLY ILLUSTRATION 6 30 ...

Page 144: ...tor Assy FRU No 100 from the Swing Arm Assy Refer to 6 2 1 3 Remove the Swing Arm Assy FRU 401 Refer to 6 2 24 4 Remove the Left Cover FRU 404 Refer to 6 2 27 5 Remove the Right Cover FRU 405 Refer to 6 2 28 6 Remove the Rear Cover FRU 408 Refer to 6 2 31 7 Remove the Pole Cover FRU 403 Refer to 6 2 26 8 Remove eleven hexagonal bolts 1 11 9 Disconnect one connector CN02 J28 10 Remove the Pipe Assy...

Page 145: ...AL SYSTEMS LOGIQ α200 SERVICE MANUAL 2138853 6 61 RENEWAL PARTS REV 5 6 2 25 Pipe Assy FRU No 402 Continued 1 2 3 4 5 6 7 8 9 10 11 402 Pipe Assy CN02 J28 Two Connectors NECK ASSY DISASSEMBLY ILLUSTRATION 6 31 ...

Page 146: ...r to ILLUSTRATION 6 32 1 Turn OFF the system 2 Remove the Monitor Assy FRU No 100 from the Swing Arm Assy Refer to 6 2 1 3 Remove the Swing Arm Assy FRU 401 Refer to 6 2 24 4 Remove the Left Cover FRU 404 Refer to 6 2 27 5 Remove the Right Cover FRU 405 Refer to 6 2 28 6 Remove the Rear Cover FRU 408 Refer to 6 2 31 7 Unscrew two screws 1and 2 8 Pull upward the Pole Cover ...

Page 147: ...GE MEDICAL SYSTEMS LOGIQ α200 SERVICE MANUAL 2138853 6 63 RENEWAL PARTS REV 5 6 2 26 Pole Cover FRU No 403 Continued 1 2 403 Pole Cover POLE COVER DISASSEMBLY ILLUSTRATION 6 32 ...

Page 148: ... 27 Left Cover FRU No 404 Time Required 5 Minutes Tool Required Screwdriver Procedure Refer to ILLUSTRATION 6 33 1 Turn OFF the system 2 Remove four screw caps and unscrew four screws 1 4 3 Remove the Left Cover FRU 404 1 2 3 4 404 Left Cover LEFT COVER DISASSEMBLY ILLUSTRATION 6 33 ...

Page 149: ...Right Cover FRU No 405 Time Required 5 Minutes Tool Required Screwdriver Procedure Refer to ILLUSTRATION 6 34 1 Turn OFF the system 2 Remove the four screw caps and unscrew four screws 1 4 3 Remove the Right Cover FRU 405 1 2 3 4 405 Right Cover RIGHT COVER DISASSEMBLY ILLUSTRATION 6 34 ...

Page 150: ... Minutes Tool Required Screwdriver Procedure Refer to ILLUSTRATION 6 35 1 Turn OFF the system 2 Remove the Left Cover FRU 404 Refer to 6 2 27 3 Remove the Right Cover FRU 405 Refer to 6 2 28 4 Unscrew two screws 1 and 2 5 Remove the Front Base Cover FRU 406 1 2 406 Front Base Cover FRONT BASE COVER DISASSEMBLY ILLUSTRATION 6 35 ...

Page 151: ...GE MEDICAL SYSTEMS LOGIQ α200 SERVICE MANUAL 2138853 6 67 RENEWAL PARTS REV 5 This page is left blank intentionally ...

Page 152: ...tes Tool Required Screwdriver Procedure Refer to ILLUSTRATION 6 36 1 Turn OFF the system 2 Remove the Left Cover FRU 404 Refer to 6 2 27 3 Remove the Right Cover FRU 405 Refer to 6 2 28 4 Remove the Front Base Cover FRU 406 Refer to 6 2 29 5 Remove the fore screw caps and unscrew fore screws 1 4 6 Remove the Front Cover FRU 407 ...

Page 153: ...GE MEDICAL SYSTEMS LOGIQ α200 SERVICE MANUAL 2138853 6 69 RENEWAL PARTS REV 5 6 2 30 Front Cover FRU No 407 Continued 1 2 407 Front Cover 3 4 FRONT COVER DISASSEMBLY ILLUSTRATION 6 36 ...

Page 154: ...ewdriver Procedure Refer to ILLUSTRATION 6 37 1 Turn OFF the system 2 Remove the Left Cover FRU 404 Refer to 6 2 27 3 Remove the Right Cover FRU 405 Refer to 6 2 28 4 Remove the four screw caps and unscrew four screws 1 4 5 Remove the Rear Cover FRU No 408 ÉÉÉÉ ÉÉÉÉ ÉÉÉÉ ÉÉÉÉ ÉÉÉÉ ÉÉÉÉ 1 2 3 4 408 Rear Cover REAR COVER DISASSEMBLY ILLUSTRATION 6 37 ...

Page 155: ...GE MEDICAL SYSTEMS LOGIQ α200 SERVICE MANUAL 2138853 6 71 RENEWAL PARTS REV 5 This page is left blank intentionally ...

Page 156: ...ver Procedure Refer to ILLUSTRATION 6 38 1 Turn OFF the system 2 Remove the Left Cover FRU 404 Refer to 6 2 27 3 Remove the Right Cover FRU 405 Refer to 6 2 28 4 Remove the Rear Cover FRU 408 Refer to 6 2 31 5 Pull up the Pole Cover FRU 403 Refer to 6 2 22 6 Remove the Cable Hook Arm 1 7 Unscrew five screws 2 3 and four bolts 4 7 8 Remove the Top Cover ...

Page 157: ...MEDICAL SYSTEMS LOGIQ α200 SERVICE MANUAL 2138853 6 73 RENEWAL PARTS REV 5 6 2 32 Top Cover FRU No 409 Continued ÉÉÉ ÉÉÉ ÉÉÉ ÉÉÉ ÉÉÉ ÉÉÉ 4 2 3 409 Top Cover 7 6 5 1 REAR COVER DISASSEMBLY ILLUSTRATION 6 38 ...

Page 158: ...Minutes Tool Required Screwdriver Procedure Refer to ILLUSTRATION 6 39 1 Turn OFF the system 2 Remove the Front Cover FRU 407 Refer to 6 2 30 3 Remove the Top Cover FRU 409 Refer to 6 2 32 4 Unscrew the six screw 1 6 5 Remove the Printer Cover FRU No 410 1 2 3 4 5 6 410 Printer Cover PRINTER COVER DISASSEMBLY ILLUSTRATION 6 39 ...

Page 159: ...GE MEDICAL SYSTEMS LOGIQ α200 SERVICE MANUAL 2138853 6 75 RENEWAL PARTS REV 5 This page is left blank intentionally ...

Page 160: ...he Top Cover FRU 409 Refer to 6 2 32 7 Remove the Printer Cover FRU 410 Refer to 6 2 33 8 Disconnect three cables connected to Back Bracket Assy FRU 412 as listed below Cut the tie wraps off if necessary a Unscrew and Disconnect two ring terminals 1 and 2 from the Terminal Block 6 Assy Refer to 6 2 46 ILLUSTRATION 6 52 b Unscrew and disconnect one ring terminals Protective Earth from the chassis c...

Page 161: ...UAL 2138853 6 77 RENEWAL PARTS REV 5 6 2 34 Printer Bracket Assy FRU No 411 Back Bracket Assy FRU No 412 Continued 2 3 4 1 412 Back Bracket Assy 411 Printer Bracket Assy PRINTER BRACKET ASSY AND BACK BRACKET ASSY DISASSEMBLY ILLUSTRATION 6 40 ...

Page 162: ... Keyboard Assy FRU 204 Refer to 6 2 10 3 Remove the Left Cover FRU 404 Refer to 6 2 27 4 Remove the Right Cover FRU 405 Refer to 6 2 28 5 Remove the Front Base Cover FRU 406 Refer to 6 2 29 6 Remove the Front Cover FRU 407 Refer to 6 2 30 7 Remove the Top Cover FRU 409 Refer to 6 2 32 8 Remove the Printer Cover FRU 410 Refer to 6 2 33 9 Remove the Neck Frame FRU 414 Refer to 6 2 36 10 Remove four ...

Page 163: ...GE MEDICAL SYSTEMS LOGIQ α200 SERVICE MANUAL 2138853 6 79 RENEWAL PARTS REV 5 6 2 35 Rear Handle FRU No 413 Continued ÉÉÉ ÉÉÉ ÉÉÉ ÉÉÉ 1 2 3 4 413 Rear Handle REAR HANDLE DISASSEMBLY ILLUSTRATION 6 41 ...

Page 164: ...system 2 Remove the Keyboard Assy FRU No 204 Refer to 6 2 10 3 Remove the Left Cover FRU 404 Refer to 6 2 27 4 Remove the Right Cover FRU 405 Refer to 6 2 28 5 Remove the Front Base Cover FRU 406 Refer to 6 2 29 6 Remove the Front Cover FRU 407 Refer to 6 2 30 7 Remove the Top Cover FRU 409 Refer to 6 2 32 8 Remove the Printer Cover FRU 410 Refer to 6 2 33 9 Remove four hexagonal bolts 1 4 10 Remo...

Page 165: ...GE MEDICAL SYSTEMS LOGIQ α200 SERVICE MANUAL 2138853 6 81 RENEWAL PARTS REV 5 6 2 36 Neck Frame FRU No 414 Continued 1 2 3 4 414 Neck Frame NECK FRAME DISASSEMBLY ILLUSTRATION 6 42 ...

Page 166: ...ve the Rear Cover FRU 408 Refer to 6 2 31 5 Remove the EMI Cover FRU 421 Refer to 6 2 39 6 Disconnect all connectors out of the Rear Panel assy If necessary cut tie wraps Connectors are listed below a Unscrew and disconnect three ring terminals 1A 2A and 5A from the Terminal Block 6 Assy Refer to 6 2 46 ILLUSTRATION 6 52 b Disconnect four connectors J2 J3 J6 and J8 from the Mother Assy For each ca...

Page 167: ...DICAL SYSTEMS LOGIQ α200 SERVICE MANUAL 2138853 6 83 RENEWAL PARTS REV 5 6 2 37 Rear Panel Assy FRU No 415 Continued ÉÉÉ ÉÉÉ ÉÉÉ ÉÉÉ 1 2 3 4 415 Rear Panel Assy REAR PANEL ASSY DISASSEMBLY ILLUSTRATION 6 43 ...

Page 168: ...OFF the power switch Refer to ILLUSTRATION 6 44 1 Turn OFF the system and unplug the power cord from the outlet 2 Remove the Left Cover FRU 404 Refer to 6 2 27 3 Remove the Right Cover FRU 405 Refer to 6 2 28 4 Remove the Front Base Cover FRU 406 Refer to 6 2 29 5 Remove the Front Cover FRU 407 Refer to 6 2 30 6 Unscrew four screws 1 4 7 Disconnect fore Terminals from the Power S W Assy 8 Remove t...

Page 169: ...GE MEDICAL SYSTEMS LOGIQ α200 SERVICE MANUAL 2138853 6 85 RENEWAL PARTS REV 5 6 2 38 Power S W Assy FRU No 416 Continued 1 2 3 4 416 Power S W Assy POWER S W ASSY DISASSEMBLY ILLUSTRATION 6 44 ...

Page 170: ...s Tool Required Screwdriver Procedure Refer to ILLUSTRATION 6 45 1 Turn OFF the system 2 Remove the Left Cover FRU 404 Refer to 6 2 27 3 Remove the Right Cover FRU 405 Refer to 6 2 28 4 Remove the Top Cover FRU 409 Refer to 6 2 32 5 Disconnect one connector CN01 6 Unscrew four screws 1 4 on the fan 7 Pull out the Fan Assy outwards ...

Page 171: ...DICAL SYSTEMS LOGIQ α200 SERVICE MANUAL 2138853 6 87 RENEWAL PARTS REV 5 6 2 39 AC Fan Assy FRU No 417 Continued ÉÉÉ ÉÉÉ ÉÉÉ ÉÉÉ ÉÉÉ ÉÉÉ 1 2 3 4 417 AC Fan Assy CN01 AC FAN ASSY DISASSEMBLY ILLUSTRATION 6 45 ...

Page 172: ...418 Time Required 15 Minutes Tool Required Hexagonal Wrench Procedure Refer to ILLUSTRATION 6 46 1 Turn OFF the system 2 Put a box below the Power Trans box 3 Unscrew four hexagonal bolts 1 4 on each Front Caster 4 Remove the Front Caster 1 2 3 4 418 Front Caster FRONT CASTER DISASSEMBLY ILLUSTRATION 6 46 ...

Page 173: ...o 419 Time Required 15 Minutes Tool Required Hexagonal Wrench Procedure Refer to ILLUSTRATION 6 47 1 Turn OFF the system 2 Put a box below the Power Trans box 3 Unscrew four hexagonal bolts 1 4 on each Rear Caster 4 Remove the Rear Caster 1 2 3 4 419 Rear Caster REAR CASTER DISASSEMBLY ILLUSTRATION 6 47 ...

Page 174: ...rench Procedure Refer to ILLUSTRATION 6 48 1 Turn OFF the system 2 Remove the Left Cover FRU 404 Refer to 6 2 27 3 Remove the Right Cover FRU 405 Refer to 6 2 28 4 Remove the Front Base Cover FRU 406 Refer to 6 2 29 5 Unscrew three hexagonal bolts 1 3 on each Bumper Set 6 Remove the Bumper Set FRU 420 1 2 3 1 2 3 420 Bumper Set BUMPER SET DISASSEMBLY ILLUSTRATION 6 48 ...

Page 175: ...RU No 421 Time Required 5 Minutes Tool Required Screwdriver Procedure Refer to ILLUSTRATION 6 49 1 Turn OFF the system 2 Remove the Left Cover FRU 404 Refer to 6 2 27 3 Unscrew eight screws 1 8 4 Remove the EMI Cover L FRU 421 1 2 5 8 6 3 4 7 210 EMI Cover L EMI COVER L DISASSEMBLY ILLUSTRATION 6 49 ...

Page 176: ...y FRU No 422 Time Required 5 Minutes Tool Required Screwdriver Procedure Refer to ILLUSTRATION 6 34 1 Turn OFF the system 2 Remove the Right Cover FRU 405 Refer to 6 2 28 3 Unscrew four screws 1 4 4 Remove the PCB Guide Assy FRU 422 1 2 3 4 422 PCB Guide Assy PCB GUIDE DISASSEMBLY ILLUSTRATION 6 50 ...

Page 177: ...GE MEDICAL SYSTEMS LOGIQ α200 SERVICE MANUAL 2138853 6 93 RENEWAL PARTS REV 5 This page is left blank intentionally ...

Page 178: ...e power switch Refer to ILLUSTRATION 6 51 1 Turn OFF the system and unplug the power cord from the outlet 2 Remove the Left Cover FRU 404 Refer to 6 2 27 3 Unscrew the two screws 1 and 2 and pull out the Terminal Block 12 Assy 4 Unscrew and Disconnect twelve ring terminals TB2 1 TB2 12 Refer to ILLUSTRATION 5 3 Chapter 5 Cut the tie wrap if necessary 5 Unscrew and Disconnect twelve ring terminals ...

Page 179: ... α200 SERVICE MANUAL 2138853 6 95 RENEWAL PARTS REV 5 6 2 45 Terminal Block 12 Assy FRU No 501 Continued ÉÉÉ ÉÉÉ ÉÉÉ ÉÉÉ ÉÉÉ ÉÉÉ 501 Terminal Block 12 Assy TB2 12 TB2 1 1 2 TERMINAL BLOCK 12 ASSY DISASSEMBLY ILLUSTRATION 6 51 ...

Page 180: ...r to ILLUSTRATION 6 52 1 Turn OFF the system and unplug the power cord from the outlet 2 Remove the Left Cover FRU 404 Refer to 6 2 27 3 Remove the Right Cover FRU 405 Refer to 6 2 28 4 Remove the Rear Cover FRU 408 Refer to 6 2 31 5 Unscrew and Disconnect six ring terminals TB3 1 TB3 6 6 Unscrew and Disconnect six ring terminals TB3 1 TB3 6 7 Unscrew the two screws 1and 2 and pull out the Termina...

Page 181: ...0 SERVICE MANUAL 2138853 6 97 RENEWAL PARTS REV 5 6 2 46 Terminal Block 6 Assy FRU No 502 Continued ÉÉÉ ÉÉÉ ÉÉÉ ÉÉÉ ÉÉÉ ÉÉÉ 1 2 502 Terminal Block 6 Assy TB3 1 TB3 6 TB3 1 TB3 6 TERMINAL BLOCK 6 ASSY DISASSEMBLY ILLUSTRATION 6 52 ...

Page 182: ...EMS LOGIQ α200 SERVICE MANUAL 2138853 6 98 RENEWAL PARTS REV 5 6 2 47 SMPS Assy FRU No 503 Time Required 10 Minutes Tool Required Screwdriver Hexagonal Wrench Key Size 4 mm Procedure Refer to 6 4 SMPS Assy REPLACEMENT ...

Page 183: ...GE MEDICAL SYSTEMS LOGIQ α200 SERVICE MANUAL 2138853 6 99 RENEWAL PARTS REV 5 This page is left blank intentionally ...

Page 184: ...e Front Base Cover FRU 406 Refer to 6 2 29 5 Remove the Rear Cover FRU 408 Refer to 6 2 31 6 Remove the SMPS Assy FRU 503 Refer to Refer to 6 4 SMPS Assy REPLACEMENT 7 Disconnect sixteen cables listed below Cut the tie wraps off if necessary For each cable refer to ILLUSTRATION 6 53 a Unscrew and Disconnect twelve ring terminals TB2 1 TB2 12 from the Terminal Block 12 Assy Refer to 6 2 45 b Unscre...

Page 185: ...Q α200 SERVICE MANUAL 2138853 6 101 RENEWAL PARTS REV 5 6 2 48 Power Trans Assy FRU No 504 Continued ÉÉÉÉ ÉÉÉÉ ÉÉÉÉ ÉÉÉÉ ÉÉÉÉ ÉÉÉÉ ÉÉÉÉ 504 Power Trans Assy TB2 12 TB2 1 1 2 3 4 POWER TRANS ASSY DISASSEMBLY ILLUSTRATION 6 53 ...

Page 186: ...UTION Do NOT unplug the power cord before OFF the power switch Refer to ILLUSTRATION 6 54 1 Turn OFF the system unplug the power cord from the outlet 2 Remove the Power Trans Assy FRU 504 Refer to 6 2 48 3 Unscrew four screws 1 4 on the front of Power Trans Assy Refer to ILLUSTRATION 6 54 4 Remove the defective Circuit Breaker 1 2 3 4 CIRCUIT BREAKER DISASSEMBLY ILLUSTRATION 6 54 ...

Page 187: ...B 250 V SB Slow Blow type HV Assy F2 500 mA SB 250 V SB Slow Blow type Monitor F1 2 A SB 125 V SB Slow Blow type 6 3 2 Replacement Procedures Concerning replacing each fuse of LOGIQ α200 refer to the following steps Note This system contains no operator serviceable parts inside Please refer servicing to qualified personnel from GE See Address of Chapter 1 Parts Required FRU PART NAME QTY PART REMA...

Page 188: ...lug the power cord from the outlet Follow the OSHA Lockout Tagout Requirements 29 CFR 1910 147 Direction 46 015902 2 Remove the Left Cover FRU 404 Refer to 6 2 27 3 Remove the Right Cover FRU 405 Refer to 6 2 28 4 Remove the Front Base Cover FRU 406 Refer to 6 2 29 5 Remove the Front Cover FRU 407 Refer to 6 2 30 6 Replace the defective fuse s with new one s For the location of fuses refer to ILLU...

Page 189: ...ents 29 CFR 1910 147 Direction 46 015902 2 Remove the Monitor Bottom Cover and Monitor Rear Cover Refer to 6 2 4 3 Replace the defective fuse with new one For the location of fuses refer to ILLUSTRATION 6 56 Make sure that only fuse with the required rated current and of same type is used for replacement 4 Re install the Monitor Bottom Cover and Monitor Rear Cover F1 2A 125V MONITOR ASSY ILLUSTRAT...

Page 190: ...cting your Ultrasound Region Engineer CAUTION Possible operational damage Failure to strictly follow ESD Electrostatic Discharge precautions during this upgrade may cause constant or intermittent operational abnormalities Strictly follow all precautions Check each step as it is completed Parts Required FRU PART NAME QTY PART REMARKS 503 SMPS ASSY 1 2148233 Time Required Approximately one 1 hours T...

Page 191: ...listed below Cut the tie wraps off if necessary For each cable refer to ILLUSTRATION 6 57 a Disconnect six connectors J10 J13 J28 and J29 from the Mother Assy b Unscrew and Disconnect two ring terminals TB3 3 and TB3 5 from the Terminal Block 6 Assy c Unscrew two screws 1 and 2 ÉÉÉ ÉÉÉ ÉÉÉ ÉÉÉ J10 J11 J13 TB3 3 TB3 5 J28 J12 1 2 J29 LEFT SIDE VIEW ILLUSTRATION 6 57 ...

Page 192: ...LLUSTRATION 6 58 WARNING PERSONAL INJURY HAZARD SMPS ASSY UNIT WEIGHTS APPROXIMATELY 3 5 KILOGRAMS 7 8 POUNDS PAY ATTENTION NOT TO BE INJURED 3 Remove the SMPS Assy and place new one on the pallet 4 Slide and move the pallet to the position just under the console Make sure that the SMPS Assy unit is placed in the correct direction 5 Screw two hexagonal screws to assemble new SMPS Assy unit to the ...

Page 193: ...onnectors For the details of cables refer to 6 4 2 B Disconnecting Cables 2 Re assemble the removal parts 3 Plug the system and power it ON 4 Perform all SMPS Assy checks and leakage current test 5 Perform functional checks to verify that the system is functioning properly Refer to Chapter 4 FUNCTIONAL CHECKS in this manual 6 This completes the replacement of the SMPS Assy Unit ...

Page 194: ...GE MEDICAL SYSTEMS LOGIQ α200 SERVICE MANUAL 2138853 6 110 RENEWAL PARTS REV 5 This page is left blank intentionally ...

Page 195: ...enance 7 3 7 2 PERIODIC MAINTENANCE PROCEDURE 7 3 7 2 1 Visual Inspection 7 3 7 2 2 Cleaning 7 4 7 2 3 Measurement 7 4 7 2 4 Battery replacement 7 4 7 2 5 Note 7 4 7 3 ELECTRICAL SAFETY TESTS 7 5 7 3 1 Outlet Test Wiring Arrangement 7 5 7 3 2 Grounding Continuity 7 6 7 3 3 Chassis Leakage Current Test 7 7 7 3 4 Probe Leakage Current Test 7 10 7 3 5 When There s Too Much Leakage Current 7 14 ...

Page 196: ...GE MEDICAL SYSTEMS LOGIQ 200 SERVICE MANUAL 2138853 PERIODIC MAINTENANCE 7 2 REV 0 This page is left blank intentionally ...

Page 197: ... hours per week 7 2 PERIODIC MAINTENANCE PROCEDURE 7 2 1 Visual Inspection 1 Check all cables for cracks nicks and abrasions in their insulation 2 Check that cable sheaths are securely clamped in cable ground connector clamps where applicable 3 Check that cables are correctly routed and secured 4 Check that there are no missing screws or fasteners and replace if necessary 5 Check that all detachab...

Page 198: ...to section 2 3 ELECTRICAL SAFETY TESTS in Chapter 2 INSTALLATION for measurement procedures 7 2 4 Battery replacement If Battery Low message is displayed on the monitor replace the Battery on MST Assy 1 Print out all user preset data for reference 2 Remove MST Assy slot Number 4 from cardcage Refer to 6 2 18 PC BOARD S in chapter 6 RENEWAL PARTS for disassembly procedures 3 Insert new Battery into...

Page 199: ...ectric Safety Analyzer DALE 600E For 220 240V Units 2113015 Leakage Current Ultrasound Kit For 100 115V and 220 240V Units 7 3 1 Outlet Test Wiring Arrangement Test all outlets in the area for proper wiring arrangement by plugging in the Dale 600 600E and noting the combination of lights that are illuminated Refer to ILLUSTRATION 7 1 AC outlet Combination of Light Condition CORRECT WIRING REVERSE ...

Page 200: ... Any Caster Wheel Support GROUND CONTINUITY TEST ILLUSTRATION 7 2 Meter Procedure Follow these steps to test the ground wire resistance 1 Turn the LOGIQ 200 unit OFF 2 Plug the unit into the meter and the meter into the tested AC wall outlet Refer to ILLUSTRATION 7 2 3 Plug the black chassis cable into the meter s CHASSIS connector and attach the chassis cable clamp to an exposed metal part of the...

Page 201: ...ric Procedure The test verifies the isolation of the power line from the chassis The testing meter is connected from accessible metal parts of the case to ground Measurements should be made with the unit ON and OFF with the power line polarity Normal and Reversed Record the highest reading of current POWER OUTLET H BLACK POLARITY REVERSING SWITCH MOMENTARY SWITCH N WHITE G GREEN LEAKAGE TEST METER...

Page 202: ...AGE CURRENT TEST ILLUSTRATION 7 4 3 Plug the black chassis cable into the meter s CHASSIS connector and attach the black chassis cable clamp to an exposed metal part of the LOGIQ 200 unit Refer to ILLUSTRATION 7 4 4 Set the tester s FUNCTION switch to CHASSIS position 5 Follow the test conditions described in the following table for every test point on the unit TABLE 7 1 CHASSIS LEAKAGE CURRENT TE...

Page 203: ...RCE LEAKAGE CURRENT COUNTRY NORMAL OPEN REVERSE OPEN CONDITION GROUND POLARITY NEUTRAL USA N A 0 3mA 0 3mA N A OTHERS 0 1mA 0 5mA 0 5mA 0 5mA CAUTION Electric Shock Hazard When the meter s ground switch is OPEN don t touch the unit TABLE 7 3 DATA SHEET FOR CHASSIS SOURCE LEAKAGE CURRENT Enter name of tested peripheral here ON NORM OPEN ON NORM CLOSED ON REV OPEN ON REV CLOSED OFF NORM OPEN OFF NOR...

Page 204: ...open and closed with power line polarity normal and reversed and with the unit Off and On For each combination the probe must be active to find the worst case condition POWER OUTLET H BLACK POLARITY REVERSING SWITCH MOMENTARY SWITCH N WHITE G GREEN LEAKAGE TEST METER CONSOLE PROBE SETUP FOR PROBE LEAKAGE CURRENT ILLUSTRATION 7 5 Note Each probe will have some amount of leakage current dependent on...

Page 205: ...mixture of 1 quart water with one or more grams of table salt mixed thoroughly will substitute CAUTION To avoid probe damage and possible electric shock do not immerse probes into any liquid beyond the level indicated in the probe users manual Do not touch the probe conductive liquid or any part of the unit under test while the ISO TEST switch is depressed Follow these steps to test each transduce...

Page 206: ...t 1 Turn the LOGIQ 200 unit OFF 2 Plug the unit into the test meter and the meter into the tested AC wall outlet 3 Plug the external probe into the meter s EXTERNAL connector 4 Set the meter s FUNCTION switch to EXTERNAL position 5 Connect the probe for test with the connector of the console 6 Apply a liberal amount of gel to the transducer s lens 7 Wrap the transducer in aluminum foil making sure...

Page 207: ...0 01mA 0 01mA 0 01mA 0 01mA OTHERS 0 01mA 0 05mA 0 05mA 0 05mA COUNTRY NORMAL OPEN REVERSE OPEN CONDITION GROUND POLARITY NEUTRAL USA 0 05mA 0 05mA 0 05mA 0 05mA OTHERS 0 1mA 0 5mA 0 5mA 0 5mA 1 Type BF Applied Part Leakage Current Limits Non conductive floating Surface and Cavity Transducers 2 Type CF Applied Part Leakage Current Limits Surgical Transducer and ECG Connections TABLE 7 5 DATA SHEET...

Page 208: ... indicate high leakage currents this should be suspected as a possible cause and the outlet wiring should be visually inspected Probe Fails Test the probe in another transducer interface connector If excessive leakage current is slot dependent inspect the Transducer Interface board for bent pins poor connections and ground continuity If the problem still has stayed with the probe replace the probe...

Page 209: ...GE MEDICAL SYSTEMS LOGIQ 200 SERVICE MANUAL 2138853 INSTALLATION FOR OPTIONS 8 1 REV 0 CHAPTER 8 INSTALLATION FOR OPTIONS TABLE OF CONTENTS SECTION TITLE PAGE 8 1 INTRODUCTION 8 3 ...

Page 210: ...GE MEDICAL SYSTEMS LOGIQ 200 SERVICE MANUAL 2138853 INSTALLATION FOR OPTIONS 8 2 REV 0 This page is left blank intentionally ...

Page 211: ...ANUAL 2138853 INSTALLATION FOR OPTIONS 8 3 REV 0 8 1 INTRODUCTION This chapter is provided to keep option installation instructions Place all option installation instructions in this chapter after completing the installation of the options ...

Page 212: ...GE MEDICAL SYSTEMS LOGIQ 200 SERVICE MANUAL 2138853 INSTALLATION FOR OPTIONS 8 4 REV 0 This page is left blank intentionally ...

Reviews: