6–2–46
Terminal Block 6 Assy (FRU No. 502)
Time Required
10 Minutes
Tool Required
Screwdriver
Procedure
CAUTION
Do NOT unplug the power cord before OFF the power switch.
Refer to ILLUSTRATION 6–52.
1.
Turn OFF the system and unplug the power cord from the outlet.
2.
Remove the Left Cover (FRU 404). Refer to 6–2–27.
3.
Remove the Right Cover(FRU 405). Refer to 6–2–28.
4.
Remove the Rear Cover (FRU 408). Refer to 6–2–31.
5.
Unscrew and Disconnect six ring terminals (TB3–1 ~TB3–6).
6.
Unscrew and Disconnect six ring terminals (TB3–1 ~TB3–6).
7.
Unscrew the two screws (1and 2) and pull out the Terminal Block 6 Assy.
8.
Remove the Terminal Block 6 Assy.