GE MEDICAL SYSTEMS
LOGIQ
α
200 SERVICE MANUAL
2138853
6–28
RENEWAL PARTS
REV 4
6–2–4
Monitor Cover Set (FRU No. 103)
Time Required
4 Minutes
Tool Required
Screwdriver
Procedure
Refer to ILLUSTRATION 6–10.
1.
Turn OFF the system.
2.
Lift and set the Monitor to the highest position by pushing down the Up/down Release Button located on the
Swing Arm Assy.
3.
Unscrew one screws (1) and remove the Monitor Bottom Cover.
4.
Unscrew four screws (2 – 5) and remove the Monitor Rear Cover.
5.
Unscrew four screws (6 – 9) and remove the Monitor Front Cover.
1
2
3
4
5
6
7
8
103 Monitor Cover Set
Monitor Front Cover
Monitor Bottom Cover
Monitor Rear Cover
9
MONITOR COVER ASSY DISASSEMBLY
ILLUSTRATION 6–10