RENEWAL PARTS
REV 5
6–3–2
Replacement Procedures (Continued)
(B) F1 on Monitor Assy
1. Turn the system power OFF. Follow the OSHA Lockout/Tagout Requirements (29 CFR 1910.147, Direction
46–015902).
2.
Remove the Monitor Bottom Cover and Monitor Rear Cover. Refer to 6–2–4.
3.
Replace the defective fuse with new one. For the location of fuses, refer to ILLUSTRATION 6–56. Make
sure that only fuse with the required rated current and of same type is used for replacement.
4.
Re–install the Monitor Bottom Cover and Monitor Rear Cover.
F1 2A 125V
MONITOR ASSY
ILLUSTRATION 6–56