Swing Arm Assy (FRU No. 401)
Time Required
0.6 Hour
Tool Required
Screwdriver
Hex Wrench
Procedure
Refer to ILLUSTRATION 6–30.
1.
Turn OFF the system and unplug the power cord from outlet.
2.
Remove the Monitor Assy (FRU No. 100) from the Swing Arm Assy. Refer to 6–2–1.
3.
Remove the Monitor Space Plate (FRU No.105). Refer to 6–2–6.
4.
Unscrew four screws (1 –4) and remove the Cover located at the bottom of the Swing Arm Assy.
5.
Unscrew one screw to remove Ground Terminal (
).
6.
Unscrew two supporters ( 5 and 6 ) and two screws ( 7 and 8 ).
7.
Remove the Guide.
8.
Unscrew one screws (10).
9.
Remove the Swing Arm Assy (FRU No.401).