GE MEDICAL SYSTEMS
LOGIQ
α
200 SERVICE MANUAL
2138853
INSTALLATION
2–3
REV 0
2–1
PREINSTALLATION
2–1–1
Introduction
This section describes various general electrical, operational, and environmental considerations that must be
considered before installing the LOGIQ
α
200 Ultrasound unit.
2–1–2
Power Line Requirements
The following power line parameters should be monitored for one week before installation. We recommend that
you use an analyzer Dranetz Model 606–3 or Dranetz Model 626:
PARAMETER
:
LIMITS
Voltage Range
: Korea
: 220 VAC
±
10% (198 – 242 VAC)
: USA
: 115 VAC
±
10% (103 – 127 VAC)
: Europe, LA
: 220 – 240 VAC
10%(198 –264 VAC)
: Japan.
: 100 VAC
±
10% (90 – 110 VAC)
: Asia
: 115 VAC
±
10% (103 – 127 VAC)
: 220 – 240 VAC
10%(198 –264 VAC)
Voltage Range
: All applications
: MAX. 500 VA
Line Frequency
: All applications
: 50/60Hz (
±
2Hz)
Power Transients
: Less than 25 % of nominal peak voltage for less than 1 millisecond for any type of
transient, including line frequency, synchronous, asynchronous, or aperiodic
transients.
Decaying Oscillation : Less than 15 % of peak voltage for less than 1 millisecond.