Pipe Assy (FRU No. 402)
Time Required
0.4 Hour
Tool Required
Screwdriver
Hex Wrench
Procedure
Refer to ILLUSTRATION 6–31.
1.
Turn OFF the system and unplug the power cord from outlet.
2.
Remove the Monitor Assy (FRU No. 100) from the Swing Arm Assy. Refer to 6–2–1.
3.
Remove the Swing Arm Assy (FRU 401). Refer to 6–2–24.
4.
Remove the Left Cover (FRU 404). Refer to 6–2–27.
5.
Remove the Right Cover(FRU 405). Refer to 6–2–28.
6.
Remove the Rear Cover (FRU 408). Refer to 6–2–31.
7.
Remove the Pole Cover (FRU 403) Refer to 6–2–26.
8.
Remove eleven hexagonal bolts (1 – 11).
9.
Disconnect one connector(CN02,J28).
10.
Remove the Pipe Assy.
NOTE
When assembling the Tilt assy, fix two screw(1,7) with one drop of Loctite262.