GE MEDICAL SYSTEMS
LOGIQ
α
200 SERVICE MANUAL
2138853
6–84
RENEWAL PARTS
REV 5
6–2–38
Power S/W Assy (FRU No. 416)
Time Required
10 Minutes
Tool Required
Screwdriver
Procedure
CAUTION
Do NOT unplug the power cord before OFF the power switch.
Refer to ILLUSTRATION 6–44.
1.
Turn OFF the system and unplug the power cord from the outlet.
2.
Remove the Left Cover (FRU 404). Refer to 6–2–27.
3.
Remove the Right Cover(FRU 405). Refer to 6–2–28.
4.
Remove the Front Base Cover (FRU 406). Refer to 6–2–29.
5.
Remove the Front Cover (FRU 407). Refer to 6–2–30.
6.
Unscrew four screws (1 – 4).
7.
Disconnect fore Terminals from the Power S/W Assy.
8.
Remove the Power S/W Assy.