GE MEDICAL SYSTEMS
LOGIQ
α
200 SERVICE MANUAL
2138853
6–36
RENEWAL PARTS
REV 4
6–2–10
Keyboard Assy (FRU No. 204)
Time Required
6 Minutes
Tool Required
Screwdriver
Procedure
Refer to ILLUSTRATION 6–16.
1.
Turn OFF the system.
2.
Unscrew two screws (1 and 2) on the front bottom of Keyboard.
3.
Unscrew three screws on the PCB that hold Ground Terminal wire and remove three clamps.
4.
Unscrew two ground screws (
) and remove two Ground Terminals .
5.
Disconnect the connector and cut the tie wraps off.
6.
Pull the Keyboard Assy out with KB Cover Top Assy.
7.
Unscrew eight screws (3 – 10) and remove KB Cover Top Assy.